1.
J Assoc Res Otolaryngol ; 23(6): 815-833, 2022 12.
Artigo em Inglês | MEDLINE | ID: mdl-36050508

RESUMO

The vestibular system is responsible for our sense of balance and spatial orientation. Recent studies have shown the possibility of partially restoring the function of this system using vestibular implants. Electrical modeling is a valuable tool in assisting the development of these implants by analyzing stimulation effects. However, previous modeling approaches of the vestibular system assumed quasi-static conditions. In this work, an extended modeling approach is presented that considers the reactive component of impedance and the electrode-tissue interface and their effects are investigated in a 3D human vestibular computer model. The Fourier finite element method was employed considering the frequency-dependent electrical properties of the different tissues. The electrode-tissue interface was integrated by an instrumental electrode model. A neuron model of myelinated fibers was employed to predict the nerve responses to the electrical stimulus. Morphological changes of the predicted voltage waveforms considering the dielectric tissue properties were found compared to quasi-static simulations, particularly during monopolar electrode configuration. Introducing the polarization capacitance and the scar tissue around the electrode in combination with a power limitation leads to a considerable current reduction applied through the active electrode and, consequently, to reduced voltage amplitudes of the stimulus waveforms. The reactive component of impedance resulted in better selectivity for the excitation of target nerves compared to the quasi-static simulation at the expense of slightly increased stimulus current amplitudes. We conclude that tissue permittivity and effects of the electrode-tissue interface should be considered to improve the accuracy of the simulations.


Assuntos
Vestíbulo do Labirinto , Humanos , Simulação por Computador , Impedância Elétrica , Estimulação Elétrica/métodos , Eletrodos
2.
Front Neurosci ; 12: 588, 2018.
Artigo em Inglês | MEDLINE | ID: mdl-30214391

RESUMO

A dysfunctional vestibular system can be a severe detriment to the quality of life of a patient. Recent studies have shown the feasibility for a vestibular implant to restore rotational sensation via electrical stimulation of vestibular ampullary nerves. However, the optimal stimulation site for selective elicitation of the desired nerve is still unknown. We realized a finite element model on the basis of µCT scans of a human inner ear and incorporated naturally distributed, artificial neural trajectories. A well-validated neuron model of myelinated fibers was incorporated to predict nerve responses to electrical stimulation. Several virtual electrodes were placed in locations of interest inside the bony labyrinth (intra-labyrinthine) and inside the temporal bone, near the target nerves (extra-labyrinthine), to determine preferred stimulation sites and electrode insertion depths. We investigated various monopolar and bipolar electrode configurations as well as different pulse waveform shapes for their ability to selectively stimulate the target nerve and for their energy consumption. The selectivity was evaluated with an objective measure of the fiber recruitment. Considerable differences of required energy and achievable selectivity between the configurations were observed. Bipolar, intra-labyrinthine electrodes provided the best selectivities but also consumed the highest amount of energy. Bipolar, extra-labyrinthine configurations did not offer any advantages compared to the monopolar approach. No selective stimulation could be performed with the monopolar, intra-labyrinthine approach. The monopolar, extra-labyrinthine electrodes required the least energy for satisfactory selectivities, making it the most promising approach for functional vestibular implants. Different pulse waveform shapes did not affect the achieved selectivity considerably but shorter pulse durations showed consistently a more selective activation of the target nerves. A cathodic, centered triangular waveform shape was identified as the most energy-efficient of the tested shapes. Based on these simulations we are able to recommend the monopolar, extra-labyrinthine stimulation approach with cathodic, centered triangular pulses as good trade-off between selectivity and energy consumption. Future implant designs could benefit from the findings presented here.

3.
Front Neurosci ; 12: 107, 2018.
Artigo em Inglês | MEDLINE | ID: mdl-29535601

RESUMO

Stable posture and body movement in humans is dictated by the precise functioning of the ampulla organs in the semi-circular canals. Statistical analysis of the interrelationship between bony and membranous compartments within the semi-circular canals is dependent on the visualization of soft tissue structures. Thirty-one human inner ears were prepared, post-fixed with osmium tetroxide and decalcified for soft tissue contrast enhancement. High resolution X-ray microtomography images at 15 µm voxel-size were manually segmented. This data served as templates for centerline generation and cross-sectional area extraction. Our estimates demonstrate the variability of individual specimens from averaged centerlines of both bony and membranous labyrinth. Centerline lengths and cross-sectional areas along these lines were identified from segmented data. Using centerlines weighted by the inverse squares of the cross-sectional areas, plane angles could be quantified. The fit planes indicate that the bony labyrinth resembles a Cartesian coordinate system more closely than the membranous labyrinth. A widening in the membranous labyrinth of the lateral semi-circular canal was observed in some of the specimens. Likewise, the cross-sectional areas in the perilymphatic spaces of the lateral canal differed from the other canals. For the first time we could precisely describe the geometry of the human membranous labyrinth based on a large sample size. Awareness of the variations in the canal geometry of the membranous and bony labyrinth would be a helpful reference in designing electrodes for future vestibular prosthesis and simulating fluid dynamics more precisely.

4.
Front Neurosci ; 11: 713, 2017.
Artigo em Inglês | MEDLINE | ID: mdl-29311790

RESUMO

Our sense of balance and spatial orientation strongly depends on the correct functionality of our vestibular system. Vestibular dysfunction can lead to blurred vision and impaired balance and spatial orientation, causing a significant decrease in quality of life. Recent studies have shown that vestibular implants offer a possible treatment for patients with vestibular dysfunction. The close proximity of the vestibular nerve bundles, the facial nerve and the cochlear nerve poses a major challenge to targeted stimulation of the vestibular system. Modeling the electrical stimulation of the vestibular system allows for an efficient analysis of stimulation scenarios previous to time and cost intensive in vivo experiments. Current models are based on animal data or CAD models of human anatomy. In this work, a (semi-)automatic modular workflow is presented for the stepwise transformation of segmented vestibular anatomy data of human vestibular specimens to an electrical model and subsequently analyzed. The steps of this workflow include (i) the transformation of labeled datasets to a tetrahedra mesh, (ii) nerve fiber anisotropy and fiber computation as a basis for neuron models, (iii) inclusion of arbitrary electrode designs, (iv) simulation of quasistationary potential distributions, and (v) analysis of stimulus waveforms on the stimulation outcome. Results obtained by the workflow based on human datasets and the average shape of a statistical model revealed a high qualitative agreement and a quantitatively comparable range compared to data from literature, respectively. Based on our workflow, a detailed analysis of intra- and extra-labyrinthine electrode configurations with various stimulation waveforms and electrode designs can be performed on patient specific anatomy, making this framework a valuable tool for current optimization questions concerning vestibular implants in humans.

6.
Cochlear Implants Int ; 15(6): 318-26, 2014 Nov.
Artigo em Inglês | MEDLINE | ID: mdl-25118009

RESUMO

OBJECTIVES: A finite element model of the human coiled cochlea was used to model the voltage distribution due to stimulation by the individual electrodes of a cochlear implant. The scalar position of the electrode array was also varied in order to investigate its effect on the voltage distribution. Multi-electrode current focusing methods were then investigated, with the aim of increasing spatial selectivity. METHODS: Simultaneous current focusing is initially achieved, as in previous publications, by calculating the input currents to the 22 electrodes that best separates the voltages at these electrode positions. The benefits of this electrode focusing strategy do not, however, entirely carry over to the predicted voltage distributions at the position of the spiral ganglion cells, where excitation is believed to occur. A novel focusing strategy is then simulated, which compensates for the impedances between the currents at the electrode sites and the voltage distribution directly at the position of the spiral ganglion cells. RESULTS: The new strategy produces much better focusing at the sites of the spiral ganglion cells, as expected, but at the cost of increased current requirements. Regularization was introduced in order to reduce current requirements, which also reduced the sensitivity of the solution to uncertainties in the impedance matrix, so that improved focusing was achieved with similar current requirements to that of electrode focusing. DISCUSSION: Although such focusing strategies cannot be achieved in practice at the moment, since the responses from the electrodes to the neural sites cannot be determined with currently available recording methods, these results do support the feasibility of a more effective focusing strategy, which may provide improved spectral resolution leading to improved perception of sound.


Assuntos
Implantes Cocleares , Estimulação Elétrica/métodos , Modelos Neurológicos , Cóclea/inervação , Cóclea/cirurgia , Nervo Coclear/fisiologia , Análise de Elementos Finitos , Humanos , Gânglio Espiral da Cóclea/fisiologia