1.
Rev Sci Instrum ; 93(8): 083901, 2022 Aug 01.
Artigo em Inglês | MEDLINE | ID: mdl-36050120

RESUMO

We present a portable CO2 laser heating system for in situ x-ray absorption spectroscopy (XAS) studies at 16-BM-D (High Pressure Collaborative Access Team, Advanced Photon Source, Argonne National Laboratory). Back scattering optical measurements are made possible by the implementation of a Ge beamsplitter. Optical pyrometry is conducted in the near-infrared, and our temperature measurements are free of chromatic aberration due to the implementation of the peak-scaling method [A. Kavner and W. R. Panero, Phys. Earth Planet. Inter. 143-144, 527-539 (2004) and A. Kavner and C. Nugent, Rev. Sci. Instrum. 79, 024902 (2008)] and mode scrambling of the input signal. Laser power stabilization is established using electronic feedback, providing a steady power over second timescales [Childs et al., Rev. Sci. Instrum. 91, 103003 (2020)]-crucial for longer XAS collections. Examples of in situ high pressure-temperature extended x-ray absorption fine structure measurements of ZrO2 are presented to demonstrate this new capability.

2.
Rev Sci Instrum ; 86(9): 093907, 2015 Sep.
Artigo em Inglês | MEDLINE | ID: mdl-26429457

RESUMO

The precision of melting curve measurements using laser-heated diamond anvil cell (LHDAC) is largely limited by the correct and reliable determination of the onset of melting. We present a novel image analysis of speckle interference patterns in the LHDAC as a way to define quantitative measures which enable an objective determination of the melting transition. Combined with our low-temperature customized IR pyrometer, designed for measurements down to 500 K, our setup allows studying the melting curve of materials with low melting temperatures, with relatively high precision. As an application, the melting curve of Te was measured up to 35 GPa. The results are found to be in good agreement with previous data obtained at pressures up to 10 GPa.

SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA