Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
Nucl Med Commun ; 37(10): 1046-52, 2016 Oct.
Artigo em Inglês | MEDLINE | ID: mdl-27218429

RESUMO

OBJECTIVE: When handling Y-labelled and Lu-labelled radiopharmaceuticals, skin exposure is mainly due to ß-particles. This study aimed to investigate the equivalent dose saving of the staff when changing from an essentially manual radiolabelling procedure to an automatic dose dispenser (ADD). MATERIALS AND METHODS: The chemist and physician were asked to wear thermoluminescence dosimeters on their fingertips to evaluate the quantity of Hp(0.07) on the skin. Data collected were divided into two groups: before introducing ADD (no ADD) and after introducing ADD. RESULTS: For the chemist, the mean values (95th percentile) of Hp(0.07) for no ADD and ADD are 0.030 (0.099) and 0.019 (0.076) mSv/GBq, respectively, for Y, and 0.022 (0.037) and 0.007 (0.023) mSv/GBq, respectively, for Lu. The reduction for ADD was significant (t-test with P<0.05) for both isotopes. The relative differences before and after ADD collected for every finger were treated using the Wilcoxon test, proving a significantly higher reduction in extremity dose to each fingertip for Lu than for Y (P<0.05). For the medical staff, the mean values of Hp(0.07) (95th percentile) for no ADD and ADD are 0.021 (0.0762) and 0.0143 (0.0565) mSv/GBq, respectively, for Y, and 0.0011 (0.00196) and 0.0009 (0.00263) mSv/GBq, respectively, for Lu. The t-test provided a P-value less than 0.05 for both isotopes, making the difference between ADD and no ADD significant. CONCLUSION: ADD positively affects the dose saving of the chemist in handling both isotopes. For the medical staff not directly involved with the introduction of the ADD system, the analysis shows a learning curve of the workers over a 5-year period. Specific devices and procedures allow staff skin dose to be limited.


Assuntos
Lutécio , Exposição Ocupacional/prevenção & controle , Doses de Radiação , Receptores de Peptídeos/uso terapêutico , Pele/efeitos da radiação , Dosimetria Termoluminescente , Radioisótopos de Ítrio , Automação , Partículas beta/uso terapêutico , Química , Dedos/efeitos da radiação , Humanos , Exposição Ocupacional/análise , Médicos , Proteção Radiológica
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...