1.
Annu Int Conf IEEE Eng Med Biol Soc ; 2021: 1226-1229, 2021 11.
Artigo em Inglês | MEDLINE | ID: mdl-34891508

RESUMO

This study optimally designed and implemented highly sensitive microscale interdigitated electrodes (IDEs) to monitor microorganisms' growth in diverse environments. Gold interdigitated electrodes (AuIDE) with 4 mm×4 mm effective sensing area and varying microscale interdigitate gaps were designed and fabricated. The electrodes were electrically characterized voltametrically. Electrochemical impedance spectroscopy (EIS) measurements were conducted to determine the optimal geometry by observing the impedance spectra of microelectrodes through varying pH and temperature. Furthermore, the sensors sensitivity was evaluated by measuring the impedance properties of a microscale volume of microorganism concentrations in growth media solution.


Assuntos
Espectroscopia Dielétrica , Ouro , Impedância Elétrica , Microeletrodos