Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros











Base de dados
Intervalo de ano de publicação
1.
Anal Chem ; 83(7): 2469-75, 2011 Apr 01.
Artigo em Inglês | MEDLINE | ID: mdl-21410136

RESUMO

The use of broad bandwidth lasers with automated feedback control of wavelength was applied to the measurement of (235)U/(238)U ratios by resonance ionization mass spectrometry (RIMS) to decrease laser-induced isotopic fractionation. By broadening the bandwidth of the first laser in a three-color, three-photon ionization process from a bandwidth of 1.8 GHz to about 10 GHz, the variation in sequential relative isotope abundance measurements decreased from 10% to less than 0.5%. This procedure was demonstrated for the direct interrogation of uranium oxide targets with essentially no sample preparation.

SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA