1.
Rev Sci Instrum ; 94(1): 013506, 2023 Jan 01.
Artigo em Inglês | MEDLINE | ID: mdl-36725564

RESUMO

Thomson scattered light is polarized in the same orientation as the incident laser beam at low electron temperatures (Te). At high Te, part of the spectrum begins to become randomly polarized due to relativistic reasons. First measurements of the depolarized Thomson scattering spectrum were obtained from Joint European Torus (JET) pulses in 2016. This paper builds upon these initial measurements with the data obtained during 2021. These new measurements improve upon first results, in particular, by obtaining spectral measurements of the depolarized spectrum. The recent JET campaign was well suited to these measurements with long and hot plasmas. The resulting data are averaged over many plasmas and laser pulses to obtain a measurement of the amount of "p" and "s" scattered light as a function of Te. This experimentally obtained d(p/s)/dTe is then fitted and found to show reasonable agreement with the theoretically predicted depolarized fraction. Error estimates on the measured "p/s" have been obtained and show that the measurements are meaningful. This is good news for ITER for which the intention is to use this measurement as a check on Te determined by the core plasma Thomson scattering diagnostic by using conventional spectral measurement techniques.

2.
Rev Sci Instrum ; 93(11): 113549, 2022 Nov 01.
Artigo em Inglês | MEDLINE | ID: mdl-36461471

RESUMO

A new Doppler backscattering (DBS) system has been installed and tested on the MAST-U spherical tokamak. It utilizes eight simultaneous fixed frequency probe beams (32.5, 35, 37.5, 40, 42.5, 45, 47.5, and 50 GHz). These frequencies provide a range of radial positions from the edge plasma to the core depending on plasma conditions. The system utilizes a combination of novel features to provide remote control of the probed density wavenumber, the launched polarization (X vs O-mode), and the angle of the launched DBS to match the magnetic field pitch angle. The range of accessible density turbulence wavenumbers (kθ) is reasonably large with normalized wavenumbers kθρs ranging from ≤0.5 to 9 (ion sound gyroradius ρs = 1 cm). This wavenumber range is relevant to a variety of instabilities believed to be important in establishing plasma transport (e.g., ion temperature gradient, trapped electron, electron temperature gradient, micro-tearing, kinetic ballooning modes). The system is specifically designed to address the requirement of density fluctuation wavevector alignment which can significantly reduce the SNR if not accounted for.

3.
Rev Sci Instrum ; 93(10): 103534, 2022 Oct 01.
Artigo em Inglês | MEDLINE | ID: mdl-36319361

RESUMO

MAST-U is equipped with a Super-X divertor, which aims to reduce heat flux to the target and promote detachment. Measurements of plasma electron density and temperature in the Super-X chamber offer insight into the processes at work in this type of divertor. First data have been obtained from the MAST-U divertor Thomson scattering diagnostic designed to measure these quantities. Following a Raman scattering calibration in nitrogen, the diagnostic operated over a number of plasma pulses in the first physics campaign. Electron density and temperature measurements have been taken in attached and detached conditions as the strike leg moved through the field of view of the diagnostic. The system operated with a dedicated 30 Hz laser with timing synchronized to seven similar lasers installed in the core Thomson system. Electron densities in the range of 1 × 1018-5 × 1019 m-3 have been measured by the system throughout these regimes. Although the system was specified to measure from 1 to 40 eV, electron temperatures in the Super-X divertor in the first campaign were low, and measurement down to 0.5 eV has been critical, particularly close to the detachment front. This generation of polychromator has been designed with increased stray light rejection compared to those used in the core system. This has proved successful with very low levels of stray light observed.

4.
Rev Sci Instrum ; 92(4): 043545, 2021 Apr 01.
Artigo em Inglês | MEDLINE | ID: mdl-34243443

RESUMO

A new divertor Thomson scattering system has been developed for the MAST-U tokamak. The diagnostic will produce electron density and temperature profiles along the Super-X strike leg. The existing polychromator design has been adapted for low temperature measurements. A new 1061 nm channel with 2 nm bandwidth has been added to enable measurements down below the previous ∼5 eV limit on the core system. The optical filters used in the system have OD6 light rejection alongside a 1064.1 nm laser line filter to reduce stray light in the digitized channels. A new averaging technique has been applied to the scattered signal traces to improve the core Thomson data in the scrape-off layer. The technique reduces the systematic noise level in this region. This leads to a reduction in the error values for electron density and temperature measurements and, in particular, the digitizer noise. The technique has been applied to produce a radial profile for a number of L-mode MAST discharges down to very low densities of ∼1 × 1018 m-3.

5.
Rev Sci Instrum ; 89(10): 10C112, 2018 Oct.
Artigo em Inglês | MEDLINE | ID: mdl-30399970

RESUMO

In the ITER Core Plasma Thomson Scattering, the scattered light collection optics system is installed both inside and outside the diagnostic port under vacuum. The length of the optical path (∼6 m) and the need to shield the neutron and γ radiation increased the complexity of the system with the inclusion of multiple dog-legs, forcing the use of many elements with optical power. Multiple rounds of design have been required in order to satisfy iteratively the system requirements in terms of resolution, aberration, and shielding. The adoption of quasi-free-form reflective surfaces for several mirrors eventually allowed the correct compromise between all conflicting requirements.

6.
Rev Sci Instrum ; 87(1): 013507, 2016 Jan.
Artigo em Inglês | MEDLINE | ID: mdl-26827321

RESUMO

The Joint European Torus (JET) high resolution Thomson scattering (HRTS) system measures radial electron temperature and density profiles. One of the key capabilities of this diagnostic is measuring the steep pressure gradient, termed the pedestal, at the edge of JET plasmas. The pedestal is susceptible to limiting instabilities, such as Edge Localised Modes (ELMs), characterised by a periodic collapse of the steep gradient region. A common method to extract the pedestal width, gradient, and height, used on numerous machines, is by performing a modified hyperbolic tangent (mtanh) fit to overlaid profiles selected from the same region of the ELM cycle. This process of overlaying profiles, termed ELM synchronisation, maximises the number of data points defining the pedestal region for a given phase of the ELM cycle. When fitting to HRTS profiles, it is necessary to incorporate the diagnostic radial instrument function, particularly important when considering the pedestal width. A deconvolved fit is determined by a forward convolution method requiring knowledge of only the instrument function and profiles. The systematic error due to the deconvolution technique incorporated into the JET pedestal fitting tool has been documented by Frassinetti et al. [Rev. Sci. Instrum. 83, 013506 (2012)]. This paper seeks to understand and quantify the systematic error introduced to the pedestal width due to ELM synchronisation. Synthetic profiles, generated with error bars and point-to-point variation characteristic of real HRTS profiles, are used to evaluate the deviation from the underlying pedestal width. We find on JET that the ELM synchronisation systematic error is negligible in comparison to the statistical error when assuming ten overlaid profiles (typical for a pre-ELM fit to HRTS profiles). This confirms that fitting a mtanh to ELM synchronised profiles is a robust and practical technique for extracting the pedestal structure.

7.
Phys Rev Lett ; 115(21): 215004, 2015 Nov 20.
Artigo em Inglês | MEDLINE | ID: mdl-26636857

RESUMO

Electron and ion heating characteristics during merging reconnection start-up on the MAST spherical tokamak have been revealed in detail using a 130 channel yttrium aluminum garnet (YAG) and a 300 channel Ruby-Thomson scattering system and a new 32 chord ion Doppler tomography diagnostic. Detailed 2D profile measurements of electron and ion temperature together with electron density have been achieved for the first time and it is found that electron temperature forms a highly localized hot spot at the X point and ion temperature globally increases downstream. For the push merging experiment when the guide field is more than 3 times the reconnecting field, a thick layer of a closed flux surface form by the reconnected field sustains the temperature profile for longer than the electron and ion energy relaxation time ~4-10 ms, both characteristic profiles finally forming a triple peak structure at the X point and downstream. An increase in the toroidal guide field results in a more peaked electron temperature profile at the X point, and also produces higher ion temperatures at this point, but the ion temperature profile in the downstream region is unaffected.

8.
Rev Sci Instrum ; 85(11): 11D701, 2014 Nov.
Artigo em Inglês | MEDLINE | ID: mdl-25430211

RESUMO

The proton detector (PD) measures 3 MeV proton yield distributions from deuterium-deuterium fusion reactions within the Mega Amp Spherical Tokamak (MAST). The PD's compact four-channel system of collimated and individually oriented silicon detectors probes different regions of the plasma, detecting protons (with gyro radii large enough to be unconfined) leaving the plasma on curved trajectories during neutral beam injection. From first PD data obtained during plasma operation in 2013, proton production rates (up to several hundred kHz and 1 ms time resolution) during sawtooth events were compared to the corresponding MAST neutron camera data. Fitted proton emission profiles in the poloidal plane demonstrate the capabilities of this new system.

9.
Rev Sci Instrum ; 85(11): 11D861, 2014 Nov.
Artigo em Inglês | MEDLINE | ID: mdl-25430274

RESUMO

A single chord two-color CO2/HeNe (10.6/0.633 µm) heterodyne laser interferometer has been designed to measure the line integral electron density along the mid-plane of the MAST Upgrade tokamak, with a typical error of 1 × 10(18) m(-3) (∼2° phase error) at 4 MHz temporal resolution. To ensure this diagnostic system can be restored from any failures without stopping MAST Upgrade operations, it has been located outside of the machine area. The final design and initial testing of this system, including details of the optics, vibration isolation, and a novel phase detection scheme are discussed in this paper.

10.
Rev Sci Instrum ; 85(11): 11E431, 2014 Nov.
Artigo em Inglês | MEDLINE | ID: mdl-25430338

RESUMO

The core Thomson scattering diagnostic (TS) on the COMPASS tokamak was put in operation and reported earlier. Implementation of edge TS, with spatial resolution along the laser beam up to ∼1/100 of the tokamak minor radius, is presented now. The procedure for spatial calibration and alignment of both core and edge systems is described. Several further upgrades of the TS system, like a triggering unit and piezo motor driven vacuum window shutter, are introduced as well. The edge TS system, together with the core TS, is now in routine operation and provides electron temperature and density profiles.

11.
Rev Sci Instrum ; 84(10): 103507, 2013 Oct.
Artigo em Inglês | MEDLINE | ID: mdl-24188274

RESUMO

This work isolated the cause of the observed discrepancy between the electron temperature (T(e)) measurements before and after the JET Core LIDAR Thomson Scattering (TS) diagnostic was upgraded. In the upgrade process, stray light filters positioned just before the detectors were removed from the system. Modelling showed that the shift imposed on the stray light filters transmission functions due to the variations in the incidence angles of the collected photons impacted plasma measurements. To correct for this identified source of error, correction factors were developed using ray tracing models for the calibration and operational states of the diagnostic. The application of these correction factors resulted in an increase in the observed T(e), resulting in the partial if not complete removal of the observed discrepancy in the measured T(e) between the JET core LIDAR TS diagnostic, High Resolution Thomson Scattering, and the Electron Cyclotron Emission diagnostics.

12.
Rev Sci Instrum ; 83(10): 10E312, 2012 Oct.
Artigo em Inglês | MEDLINE | ID: mdl-23126972

RESUMO

A real-time system has been developed to trigger both the MAST Thomson scattering (TS) system and the plasma control system on the phase and amplitude of neoclassical tearing modes (NTMs), extending the capabilities of the original system. This triggering system determines the phase and amplitude of a given NTM using magnetic coils at different toroidal locations. Real-time processing of the raw magnetic data occurs on a low cost field programmable gate array (FPGA) based unit which permits triggering of the TS lasers on specific amplitudes and phases of NTM evolution. The MAST plasma control system can receive a separate trigger from the FPGA unit that initiates a vertical shift of the MAST magnetic axis. Such shifts have fully removed m∕n = 2∕1 NTMs instabilities on a number of MAST discharges.

13.
Rev Sci Instrum ; 83(10): 10E350, 2012 Oct.
Artigo em Inglês | MEDLINE | ID: mdl-23127007

RESUMO

The electron density and temperature profiles measured by the Thomson scattering diagnostic on the COMPASS tokamak are used for estimation of electron kinetic energy, energy confinement time, and effective charge number Z(eff). Data are compared with the line-integrated electron density measured by a microwave interferometer in an ohmically heated plasma with a circular cross section. An error analysis of both electron temperature and density are performed by two methods-a constant chi-square boundaries method and a Monte Carlo simulation, determining asymmetrical error bars for the electron temperature.

14.
Phys Rev Lett ; 108(13): 135002, 2012 Mar 30.
Artigo em Inglês | MEDLINE | ID: mdl-22540707

RESUMO

Plasma equilibria reconstructed from the Mega-Amp Spherical Tokamak have sufficient resolution to capture plasma evolution during the short period between edge-localized modes (ELMs). Immediately after the ELM, steep gradients in pressure, P, and density, n(e), form pedestals close to the separatrix, and they then expand into the core. Local gyrokinetic analysis over the ELM cycle reveals the dominant microinstabilities at perpendicular wavelengths of the order of the ion Larmor radius. These are kinetic ballooning modes in the pedestal and microtearing modes in the core close to the pedestal top. The evolving growth rate spectra, supported by gyrokinetic analysis using artificial local equilibrium scans, suggest a new physical picture for the formation and arrest of this pedestal.

15.
Rev Sci Instrum ; 83(1): 013506, 2012 Jan.
Artigo em Inglês | MEDLINE | ID: mdl-22299950

RESUMO

The instrument function of the high resolution Thomson scattering (HRTS) diagnostic in the Joint European Torus (JET) has been calculated for use in improved pedestal profile analysis. The full width at half maximum (FWHM) of the spatial instrument response is (22 ± 1) mm for the original HRTS system configuration and depends on the particular magnetic topology of the JET plasmas. An improvement to the optical design of the laser input system is presented. The spatial smearing across magnetic flux surfaces is reduced in this design. The new input system has been implemented (from JPN 78742, July 2009) and the HRTS instrument function corresponding to the new configuration has been improved to approximately FWHM = (9.8 ± 0.8) mm. The reconstructed instrument kernels are used in combination with an ad hoc forward deconvolution procedure for pedestal analysis. This procedure produces good results for both the old and new setups, but the reliability of the deconvolved profiles is greatly reduced when the pedestal width is of the same order as, or less than the FWHM of the instrument kernel.

16.
Rev Sci Instrum ; 82(5): 053501, 2011 May.
Artigo em Inglês | MEDLINE | ID: mdl-21639498

RESUMO

Deconvolution of Thomson scattering (TS) profiles is required when the gradient length of the electron temperature (T(e)) or density (n(e)) are comparable to the instrument function length (Δ(R)). The most correct method for deconvolution to obtain underlying T(e) and n(e) profiles is by consideration of scattered signals. However, deconvolution at the scattered signal level is complex since it requires knowledge of all spectral and absolute calibration data. In this paper a simple technique is presented where only knowledge of the instrument function I(r) and the measured profiles, T(e, observed)(r) and n(e, observed)(r), are required to obtain underlying T(e)(r) and n(e)(r). This method is appropriate for most TS systems and is particularly important where high spatial sampling is obtained relative to Δ(R).

17.
Rev Sci Instrum ; 82(3): 033514, 2011 Mar.
Artigo em Inglês | MEDLINE | ID: mdl-21585113

RESUMO

The present work is motivated by a long standing discrepancy between the electron temperature measurements of Thomson scattering (TS) and electron cyclotron emission (ECE) diagnostics for plasmas with strong auxiliary heating observed at both JET and TFTR above 6­7 keV, where in some cases the ECE electron temperature measurements can be 15%­20% higher than the TS measurements. Recent analysis based on ECE results at JET has shown evidence of distortions to the Maxwellian electron velocity distribution and a correlation with the TS and ECE discrepancies has been suggested. In this paper, a technique to determine the presence of non-Maxwellian behavior using TS diagnostics is outlined. The difficulties and limitations of modern TS system designs to determine the electron velocity distribution are also discussed. It is demonstrated that small deviations such as those suggested by previous ECE analysis could be potentially detected, depending on the spectral layout of the TS polychromators. The spectral layout of the JET high resolution Thomson scattering system is such that it could be used to determine these deviations between 1 and 6 keV, and the results presented here indicate that no evidence of non-Maxwellian behavior is observed in this range. In this paper, a modification to the current polychromator design is proposed, allowing non-Maxwellian distortions to be detected up to at least 10 keV.

18.
Rev Sci Instrum ; 81(10): 10D520, 2010 Oct.
Artigo em Inglês | MEDLINE | ID: mdl-21033875

RESUMO

A Thomson scattering diagnostic designed to measure both edge and core physics has been implemented on MAST. The system uses eight Nd:YAG lasers, each with a repetition rate of 30 Hz. The relative and absolute timing of the lasers may be set arbitrarily to produce fast bursts of measurements to suit the time evolution of the physics being studied. The scattered light is collected at F/6 by a 100 kg six element lens system with an aperture stop of 290 mm. The collected light is then transferred to 130 polychromators by 130 independent fiber bundles. The data acquisition and processing are based on a distributed computer system of dual core processors embedded in 26 chassis. Each chassis is standalone and performs data acquisition and processing for five polychromators. This system allows data to be available quickly after the MAST shot and has potential for real-time operations.

19.
Rev Sci Instrum ; 81(10): 10D521, 2010 Oct.
Artigo em Inglês | MEDLINE | ID: mdl-21033876

RESUMO

The recent upgrade to the MAST YAG Thomson scattering while enhancing the diagnostic capabilities increased the complexity of the system. There are eight YAG lasers now operational, doubling the number from the previous setup. This means alignment between each laser individually and reference points is essential to guarantee data quality and diagnostic reliability. To address this issue an alignment system was recently installed. It mimics the beams alignment in MAST by sampling 1% of the laser beam that is sent into a telescope which demagnifies by a factor of 8. The demagnified beam is viewed with a CCD camera. By scanning the camera the profile and position of the beams in the scattering zone and in a range of several meters inside MAST can be determined. Therefore alignment is checked along the beam path without having to sample it inside the vessel. The experimental apparatus and test procedures are described.

20.
Rev Sci Instrum ; 81(10): 10D531, 2010 Oct.
Artigo em Inglês | MEDLINE | ID: mdl-21033886

RESUMO

A new Thomson scattering diagnostic system has been designed and is being built now on the COMPASS tokamak at the Institute of Plasma Physics ASCR in Prague (IPP Prague) in the Czech Republic. This contribution focuses on design, development, and installation of the light collection and detection system. High spatial resolution of 3 mm will be achieved by a combination of design of collection optics and connected polychromators. Imaging characteristics of both core and edge plasma collection objectives are described and fiber backplane design is presented. Several calibration procedures are discussed. The operational deployment of the Thomson scattering diagnostic is planned by the end of 2010.