1.
Med Image Comput Comput Assist Interv ; 12(Pt 1): 467-74, 2009.
Artigo em Inglês | MEDLINE | ID: mdl-20426021

RESUMO

This paper describes the first accomplishment of the Time-of-Flight (ToF) measurement principle via endoscope optics. The applicability of the approach is verified by in-vitro experiments. Off-the-shelf ToF camera sensors enable the per-pixel, on-chip, real-time, marker-less acquisition of distance information. The transfer of the emerging ToF measurement technique to endoscope optics is the basis for a new generation of ToF rigid or flexible 3-D endoscopes. No modification of the endoscope optic itself is necessary as only an enhancement of illumination unit and image sensors is necessary. The major contribution of this paper is threefold: First, the accomplishment of the ToF measurement principle via endoscope optics; second, the development and validation of a complete calibration and post-processing routine; third, accomplishment of extensive in-vitro experiments. Currently, a depth measurement precision of 0.89 mm at 20 fps with 3072 3-D points is achieved.


Assuntos
Endoscópios , Aumento da Imagem/instrumentação , Imageamento Tridimensional/instrumentação , Iluminação/instrumentação , Processamento de Sinais Assistido por Computador/instrumentação , Transdutores , Desenho de Equipamento , Análise de Falha de Equipamento , Reprodutibilidade dos Testes , Sensibilidade e Especificidade