Eur J Echocardiogr ; 10(3): 389-94, 2009 May.
Artigo em Inglês | MEDLINE | ID: mdl-18945726

RESUMO

AIMS: The objective was to evaluate the function of ultrasound transducers in use in routine clinical practice and thereby estimating the incidence of defective transducers. METHODS AND RESULTS: The study comprised a one-time test of 676 transducers from 7 manufacturers which were in daily use in clinical departments at 32 hospitals. They were tested with the Sonora FirstCall Test System; 39.8% exhibited a transducer error. Delamination was detected in 26.5% and break in the cable was detected in 8.4% of the tested transducers. Errors originating from the piezoelectrical elements were unusual. Delamination and short circuit occurred without significant differences between transducers from all tested manufacturers, but the errors break in the cable, weak and dead element showed a statistically significant higher frequency in transducers from certain manufacturers. CONCLUSION: The high error frequency and the risk for incorrect medical decisions when using a defective transducer indicate an urgent need for increased testing of the transducers in clinical departments.


Assuntos
Análise de Falha de Equipamento , Transdutores/normas , Ultrassonografia Doppler em Cores/instrumentação , Criança , Permeabilidade do Canal Arterial/diagnóstico por imagem , Análise de Falha de Equipamento/métodos , Feminino , Humanos , Garantia da Qualidade dos Cuidados de Saúde , Sensibilidade e Especificidade