1.
Rev Sci Instrum
; 85(9): 095116, 2014 Sep.
Artigo
em Inglês
| MEDLINE
| ID: mdl-25273781
RESUMO
We report the design, construction, and characterization of a biaxial sample rotation stage for use in a cryogenic system for orientation-dependent studies of anisotropic electronic transport phenomena at low temperatures and high magnetic fields. Our apparatus allows for continuous rotation of a sample about two axes, both independently and simultaneously.