Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 4 de 4
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
Artigo em Inglês | MEDLINE | ID: mdl-20378459

RESUMO

The radiation impedance of a capacitive micromachined ultrasonic transducer (cMUT) with a circular membrane is calculated analytically using its velocity profile for the frequencies up to its parallel resonance frequency for both the immersion and the airborne applications. The results are verified by finite element simulations. The work is extended to calculate the radiation impedance of an array of cMUT cells positioned in a hexagonal pattern. A higher radiation resistance improves the bandwidth as well as the efficiency of the cMUT. The radiation resistance is determined to be a strong function of the cell spacing. It is shown that a center-to-center cell spacing of 1.25 wavelengths maximizes the radiation resistance, if the membranes are not too thin. It is also found that excitation of nonsymmetric modes may reduce the radiation resistance in immersion applications.

2.
Artigo em Inglês | MEDLINE | ID: mdl-20178910

RESUMO

Finite element method (FEM) is used for transient dynamic analysis of capacitive micromachined ultrasonic transducers (CMUT) and is particularly useful when the membranes are driven in the nonlinear regime. One major disadvantage of FEM is the excessive time required for simulation. Harmonic balance (HB) analysis, on the other hand, provides an accurate estimate of the steady-state response of nonlinear circuits very quickly. It is common to use Mason's equivalent circuit to model the mechanical section of CMUT. However, it is not appropriate to terminate Mason's mechanical LC section by a rigid piston's radiation impedance, especially for an immersed CMUT. We studied the membrane behavior using a transient FEM analysis and found out that for a wide range of harmonics around the series resonance, the membrane displacement can be modeled as a clamped radiator. We considered the root mean square of the velocity distribution on the membrane surface as the circuit variable rather than the average velocity. With this definition, the kinetic energy of the membrane mass is the same as that in the model. We derived the force and current equations for a clamped radiator and implemented them using a commercial HB simulator. We observed much better agreement between FEM and the proposed equivalent model, compared with the conventional model.


Assuntos
Dinâmica não Linear , Transdutores , Ultrassom , Análise de Elementos Finitos , Microtecnologia/instrumentação
3.
Artigo em Inglês | MEDLINE | ID: mdl-17571821

RESUMO

We present a lumped element parametric model for the clamped circular membrane of a capacitive micromachined ultrasonic transducer (cMUT). The model incorporates an electrical port and two sets of acoustic ports, through which the cMUT couples to the medium. The modeling approach is based on matching a lumped element model and the mechanical impedance of the cMUT membrane at the resonance frequencies in vacuum. Very good agreement between finite element simulation results and model impedance is obtained. Equivalent circuit model parameters can be found from material properties and membrane dimensions without a need for finite element simulation.


Assuntos
Transdutores , Ultrassonografia/instrumentação , Simulação por Computador , Desenho Assistido por Computador , Capacitância Elétrica , Desenho de Equipamento , Análise de Falha de Equipamento , Modelos Lineares , Membranas Artificiais , Miniaturização , Reprodutibilidade dos Testes , Sensibilidade e Especificidade , Vácuo
4.
Artigo em Inglês | MEDLINE | ID: mdl-16463487

RESUMO

Capacitive micromachined ultrasonic transducers (cMUT) have large bandwidths, but they typically have low conversion efficiencies. This paper defines a performance measure in the form of a gain-bandwidth product and investigates the conditions in which this performance measure is maximized. A Mason model corrected with finite-element simulations is used for the purpose of optimizing parameters. There are different performance measures for transducers operating in transmit, receive, or pulse-echo modes. Basic parameters of the transducer are optimized for those operating modes. Optimized values for a cMUT with silicon nitride membrane and immersed in water are given. The effect of including an electrical matching network is considered. In particular, the effect of a shunt inductor in the gain-bandwidth product is investigated. Design tools are introduced, which are used to determine optimal dimensions of cMUTs with the specified frequency or gain response.


Assuntos
Desenho Assistido por Computador , Modelos Lineares , Membranas Artificiais , Transdutores , Ultrassonografia/instrumentação , Simulação por Computador , Capacitância Elétrica , Eletrônica Médica , Desenho de Equipamento , Análise de Falha de Equipamento , Análise de Elementos Finitos , Miniaturização , Controle de Qualidade , Ultrassonografia/métodos
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...