Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
Biosens Bioelectron ; 8(3-4): 185-9, 1993.
Artigo em Inglês | MEDLINE | ID: mdl-8357572

RESUMO

In this paper the development of a surface plasmon resonance (SPR) immunosensor for syphilis screening is described. This immunosensor is based on the detection of antibodies in serum against the causative organism Treponema pallidum. In order to achieve selectivity a recombinant Treponema pallidum membrane protein A (TmpA) was used. This antigen can react with antibodies to T. pallidum, present in serum of syphilitic patients. Reproducible results have been obtained, using a 'sandwich SPR' method: binding of a sandwich antibody to the treponemal antibody after serum incubation was measured in real time while the binding was taking place. The SPR results obtained from ten blind-coded sera corresponded well with classical syphilis tests (Treponema pallidum haemagglutination assay (TPHA), fluorescent treponemal antibody-absorbed test (FTA-ABS), venereal diseases research laboratory flocculation test (VDRL) and TmpA-based enzyme-linked immunosorbent assay (TmpA-ELISA)). Preliminary experiments showed that direct measurement of serum (in the 'one step SPR') is not yet possible, probably as a result of non-uniformity of serum samples. The application of latex beads is considered to solve this problem.


Assuntos
Anticorpos Antibacterianos/sangue , Proteínas de Bactérias/imunologia , Técnicas Biossensoriais , Bancos de Sangue/normas , Serina Endopeptidases , Sífilis/prevenção & controle , Treponema pallidum/imunologia , Antígenos de Bactérias/imunologia , Ensaio de Imunoadsorção Enzimática , Desenho de Equipamento , Testes de Floculação , Testes de Hemaglutinação , Humanos , Programas de Rastreamento/métodos , Proteínas Recombinantes/imunologia , Sífilis/diagnóstico , Sífilis/transmissão
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...