1.
J Chromatogr A ; 1172(2): 151-9, 2007 Nov 23.
Artigo em Inglês | MEDLINE | ID: mdl-17961580

RESUMO

A novel system for performing high-speed counter-current chromatography has been developed for separation of biopolymers using polymer phase systems. The spiral disk assembly consisting of eight units, each equipped with over 300 mixer-settler sets, was constructed and performance evaluated in terms of retention of the stationary phase and separation efficiency. A series of experiments was performed with a polymer phase system composed of polyethylene glycol 1000 (12.5%, w/w) and dibasic potassium phosphate (12.5%, w/w) using two stable protein samples of myoglobin and lysozyme at various experimental conditions of flow rates and revolution speeds. The best results were obtained with revolution speeds of 800-1000rpm at flow rates of 0.25-0.5ml/min where the partition efficiency of several 100 theoretical plates was achieved with over 50% stationary phase retention.


Assuntos
Algoritmos , Distribuição Contracorrente/instrumentação , Distribuição Contracorrente/métodos , Polietilenoglicóis/química , Centrifugação , Cromatografia Líquida/métodos , Desenho de Equipamento , Muramidase/análise , Mioglobina/análise , Polímeros/química , Solventes/química , Água