Sci Total Environ ; 74: 249-62, 1988 Aug 01.
Artigo em Inglês | MEDLINE | ID: mdl-3222695

RESUMO

Serum calcium, magnesium, copper and zinc concentrations obtained from the analysis of 77 serum samples by inductively coupled plasma-atomic emission spectrometry (ICP-AES) are compared with the results obtained using atomic absorption spectroscopy (AAS). Similarly, serum sodium and potassium levels from the analysis of the same samples by ICP-AES are compared with the levels obtained by flame photometry. For each metal, we compare the results from both methods with a linear regression program that assumes error in both variables. The regression analysis shows that the ICP-AES method gives slightly higher calcium, copper, and zinc results and lower magnesium results than the AAS methods, and lower sodium and potassium results than the flame photometry method. Except for sodium, the correlation (r) between the results is very high (greater than or equal to 0.958), indicating that the ICP-AES results could be corrected to be equivalent to the atomic absorption or flame photometry results. The ICP-AES has the advantage of requiring less preparation and analysis time, and additional elements could be determined simultaneously in the same sample.


Assuntos
Cálcio/sangue , Cobre/sangue , Magnésio/sangue , Potássio/sangue , Sódio/sangue , Zinco/sangue , Coleta de Amostras Sanguíneas , Humanos , Indicadores e Reagentes , Espectrofotometria Atômica/métodos , Análise Espectral/métodos