Sensors (Basel) ; 11(2): 1784-93, 2011.
Artigo em Inglês | MEDLINE | ID: mdl-22319381

RESUMO

To measure precise blood pressure (BP) and pulse rate without using a cuff, we have developed an arterial pulsimeter consisting of a small, portable apparatus incorporating a Hall device. Regression analysis of the pulse wave measured during testing of the arterial pulsimeter was conducted using two equations of the BP algorithm. The estimated values of BP obtained by the cuffless arterial pulsimeter over 5 s were compared with values obtained using electronic or liquid mercury BP meters. The standard deviation between the estimated values and the measured values for systolic and diastolic BP were 8.3 and 4.9, respectively, which are close to the range of values of the BP International Standard. Detailed analysis of the pulse wave measured by the cuffless radial artery pulsimeter by detecting changes in the magnetic field can be used to develop a new diagnostic algorithm for BP, which can be applied to new medical apparatus such as the radial artery pulsimeter.


Assuntos
Artérias/fisiologia , Determinação da Pressão Arterial/instrumentação , Pressão Sanguínea/fisiologia , Magnetismo/instrumentação , Pulso Arterial/instrumentação , Algoritmos , Eletrônica , Humanos , Rádio (Anatomia)/fisiologia