1.
Zhongguo Yi Liao Qi Xie Za Zhi ; 35(6): 446-8, 2011 Nov.
Artigo em Chinês | MEDLINE | ID: mdl-22379775

RESUMO

Through five experiments, such as "Circle", "square", "side that does not shake", "burning edge", "filter", this paper discussed the effect of different experimental methods of polypropylene fiber liquid filtration membrane particles shedding. The results show that: the particles falling off the edge of the polypropylene fiber Double-sided liquid filtration membrane are very important and must be considered off the evaluation of particle pollution.


Assuntos
Filtração/instrumentação , Filtração/métodos , Falha de Equipamento , Tamanho da Partícula , Polipropilenos
2.
Zhongguo Yi Liao Qi Xie Za Zhi ; 34(2): 126-8, 2010 Mar.
Artigo em Chinês | MEDLINE | ID: mdl-20540298

RESUMO

Compare and analyze the standards related to centrifuge apparatus and make corresponding suggestions in allusion to the problems existed in the test method and physical performance during product standard drafting process.


Assuntos
Remoção de Componentes Sanguíneos/instrumentação , Centrifugação/instrumentação , Centrifugação/normas , Aprovação de Equipamentos/normas
3.
Zhongguo Yi Liao Qi Xie Za Zhi ; 34(6): 445-7, 2010 Nov.
Artigo em Chinês | MEDLINE | ID: mdl-21360986

RESUMO

A comparison test on the number of puncture coring is conducted using the normal needle and the Huber needle in the injection area of the implantable drug-supplying device separately. The result indicates that the number of coring using the Huber needle is much less than that using the normal needle during the puncturing. So it is suggested to popularize the Huber needle in the drug transfusion of the implantable drug-supplying device.


Assuntos
Bombas de Infusão Implantáveis , Agulhas , Bombas de Infusão Implantáveis/classificação
4.
Zhongguo Yi Liao Qi Xie Za Zhi ; 33(5): 374-8, 2009 Sep.
Artigo em Chinês | MEDLINE | ID: mdl-20073249

RESUMO

The major idea of this article is to discuss standardization and normalization for the product standard of medical devices. Analyze the problem related to the physical performance requirements and test methods during product standard drafting process and make corresponding suggestions.


Assuntos
Aprovação de Equipamentos , Equipamentos e Provisões/normas , Teste de Materiais/métodos , Pesos e Medidas
5.
Zhongguo Yi Liao Qi Xie Za Zhi ; 32(4): 284-6, 311, 2008 Jul.
Artigo em Chinês | MEDLINE | ID: mdl-18973040

RESUMO

This article reviews the recent progress of the finite element analysis (FEA) researches focusing on various expansion types, fatigue durability and the flexibility of the stents.


Assuntos
Análise de Elementos Finitos , Desenho de Prótese , Stents , Análise de Falha de Equipamento