In the intensive care unit, or during anesthesia, patients are attached to monitors by cables. These cables obstruct nursing staff and hinder the patients from moving freely in the hospital. However, rapidly developing wireless technologies are expected to solve these problems. To this end, this study revealed problem areas in current patient monitoring and established the most important medical parameters to monitor. In addition, usable wireless techniques for short-range data transmission were explored and currently employed wireless applications in the hospital environment were studied. The most important parameters measured of the patient include blood pressures, electrocardiography, respiration rate, heart rate and temperature. Currently used wireless techniques in hospitals are based on the WMTS and WLAN standards. There are no viable solutions for short-range data transmission from patient sensors to patient monitors, but potentially usable techniques in the future are based on the WPAN standards. These techniques include Bluetooth, ZigBee and UWB. Other suitable techniques might be based on capacitive or inductive coupling. The establishing of wireless techniques depends on ensuring the reliability of data transmission, eliminating disturbance by other wireless devices, ensuring patient data security and patient safety, and lowering the power consumption and price.