1.
J Urol ; 143(5): 909-11, 1990 May.
Artigo em Inglês | MEDLINE | ID: mdl-2184254

RESUMO

Urethral catheterization, the standard method for measurement of bladder volume, is associated with patient discomfort plus the risks of urethral trauma and urinary tract infection. A portable ultrasound instrument that automatically determines bladder volume was used for 164 determinations and the digital readout correlated with the catheterized volume (r2 equals 0.79). This instrument is a noninvasive alternative to urethral catheterization for the determination of bladder volume in most patients.


Assuntos
Ultrassonografia/instrumentação , Bexiga Urinária/anatomia & histologia , Estudos de Avaliação como Assunto , Humanos , Ultrassonografia/métodos , Cateterismo Urinário