Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 5 de 5
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
IEEE Trans Biomed Circuits Syst ; 12(2): 437-445, 2018 04.
Artigo em Inglês | MEDLINE | ID: mdl-29570069

RESUMO

This paper presents a novel two-wired active electrode that achieves ultrahigh input impedance using power supply bootstrapping. The proposed circuit reduces the input capacitance of a buffer amplifier while enabling measurements using leads with only two wires, providing a low-complexity and low-cost solution for interference rejection and artifact reduction in dc-coupled dry-contact biopotential measurements. An implemented prototype shows that, even using standard operational amplifiers, an input capacitance as low as 71 fF can be obtained, maintaining a high impedance in a 0-1 kHz bandwidth, sufficient for ECG, EEG, and EMG measurements. The circuit has a simple and easily replicable implementation that requires no individual adjustment. A common mode rejection ratio (CMRR) above 103 dB at 50 Hz was achieved and the increased rejection to interference due to the potential divider effect was experimentally tested maintaining a 92-dB CMRR at 50 Hz with a 1.2-M source impedance unbalance. ECG measurements were conducted to validate the active electrode against a traditional alternative, and a test with dry-contact EEG electrodes was successfully conducted. Although the proposed circuit is intended to acquire superficial electrophysiological signals using dry electrodes, it can be used for measurement from other high-impedance sources, such as micropipette electrodes.


Assuntos
Eletrodos , Eletrodiagnóstico/instrumentação , Processamento de Sinais Assistido por Computador/instrumentação , Amplificadores Eletrônicos , Capacitância Elétrica , Desenho de Equipamento , Humanos
2.
IEEE Trans Biomed Circuits Syst ; 10(3): 787-95, 2016 Jun.
Artigo em Inglês | MEDLINE | ID: mdl-26841414

RESUMO

In this paper we present an analysis of the voltage amplifier needed for double differential (DD) sEMG measurements and a novel, very simple circuit for implementing DD active electrodes. The three-input amplifier that standalone DD active electrodes require is inherently different from a differential amplifier, and general knowledge about its design is scarce in the literature. First, the figures of merit of the amplifier are defined through a decomposition of its input signal into three orthogonal modes. This analysis reveals a mode containing EMG crosstalk components that the DD electrode should reject. Then, the effect of finite input impedance is analyzed. Because there are three terminals, minimum bounds for interference rejection ratios due to electrode and input impedance unbalances with two degrees of freedom are obtained. Finally, a novel circuit design is presented, including only a quadruple operational amplifier and a few passive components. This design is nearly as simple as the branched electrode and much simpler than the three instrumentation amplifier design, while providing robust EMG crosstalk rejection and better input impedance using unity gain buffers for each electrode input. The interference rejection limits of this input stage are analyzed. An easily replicable implementation of the proposed circuit is described, together with a parameter design guideline to adjust it to specific needs. The electrode is compared with the established alternatives, and sample sEMG signals are obtained, acquired on different body locations with dry contacts, successfully rejecting interference sources.


Assuntos
Eletromiografia/instrumentação , Eletrônica Médica/métodos , Amplificadores Eletrônicos , Eletrodos , Eletrônica Médica/instrumentação , Humanos
3.
IEEE Trans Biomed Circuits Syst ; 6(6): 614-8, 2012 Dec.
Artigo em Inglês | MEDLINE | ID: mdl-23853262

RESUMO

Single ended (SE) amplifiers allow implementing biopotential front-ends with a reduced number of parts, being well suited for preamplified electrodes or compact EEG headboxes. On the other hand, given that each channel has independent gain; mismatching between these gains results in poor common-mode rejection ratios (CMRRs) (about 30 dB considering 1% tolerance components). This work proposes a scheme for multichannel EEG acquisition systems based on SE amplifiers and a novel digital driven right leg (DDRL) circuit, which overcome the poor CMRR of the front-end stage providing a high common mode reduction at power line frequency (up to 80 dB). A functional prototype was built and tested showing the feasibility of the proposed technique. It provided EEG records with negligible power line interference, even in very aggressive EMI environments.


Assuntos
Eletroencefalografia/instrumentação , Amplificadores Eletrônicos , Engenharia Biomédica , Fontes de Energia Elétrica , Eletrodos , Eletroencefalografia/estatística & dados numéricos , Desenho de Equipamento , Humanos , Processamento de Sinais Assistido por Computador
4.
IEEE Trans Biomed Eng ; 51(8): 1444-8, 2004 Aug.
Artigo em Inglês | MEDLINE | ID: mdl-15311831

RESUMO

Fully differential amplifiers yield large differential gains and also high common mode rejection ratio (CMRR), provided they do not include any unmatched grounded component. In biopotential measurements, however, the admissible gain of amplification stages located before dc suppression is usually limited by electrode offset voltage, which can saturate amplifier outputs. The standard solution is to first convert the differential input voltage to a single-ended voltage and then implement any other required functions, such as dc suppression and dc level restoring. This approach, however, yields a limited CMRR and may result in a relatively large equivalent input noise. This paper describes a novel fully differential biopotential amplifier based on a fully differential dc-suppression circuit that does not rely on any matched passive components, yet provides large CMRR and fast recovery from dc level transients. The proposed solution is particularly convenient for low supply voltage systems. An example implementation, based on standard low-power op amps and a single 5-V power supply, accepts input offset voltages up to +/-500 mV, yields a CMRR of 102 dB at 50 Hz, and provides, in accordance with the AAMI EC38 standard, a reset behavior for recovering from overloads or artifacts.


Assuntos
Amplificadores Eletrônicos , Eletrofisiologia/instrumentação , Análise de Falha de Equipamento , Desenho Assistido por Computador , Desenho de Equipamento , Reprodutibilidade dos Testes , Sensibilidade e Especificidade
5.
IEEE Trans Biomed Eng ; 50(3): 391-5, 2003 Mar.
Artigo em Inglês | MEDLINE | ID: mdl-12669996

RESUMO

AC coupling is essential in biopotential measurements. Electrode offset potentials can be several orders of magnitude larger than the amplitudes of the biological signals of interest, thus limiting the admissible gain of a dc-coupled front end to prevent amplifier saturation. A high-gain input stage needs ac input coupling. This can be achieved by series capacitors, but in order to provide a bias path, grounded resistors are usually included, which degrade the common mode rejection ratio (CMRR). This paper proposes a novel balanced input ac-coupling network that provides a bias path without any connection to ground, thus resulting in a high CMRR. The circuit being passive, it does not limit the differential dc input voltage. Furthermore, differential signals are ac coupled, whereas common-mode voltages are dc coupled, thus allowing the closed-loop control of the dc common mode voltage by means of a driven-right-leg circuit. This makes the circuit compatible with common-mode dc shifting strategies intended for single-supply biopotential amplifiers. The proposed circuit allows the implementation of high-gain biopotential amplifiers with a reduced number of parts, thus resulting in low power consumption. An electrocardiogram amplifier built according to the proposed design achieves a CMRR of 123 dB at 50 Hz.


Assuntos
Amplificadores Eletrônicos , Eletrocardiografia/instrumentação , Eletrofisiologia/instrumentação , Desenho de Equipamento , Análise de Falha de Equipamento , Controle de Qualidade
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...