1.
Trans Am Soc Artif Intern Organs
; 24: 341-4, 1978.
Artigo
em Inglês
| MEDLINE
| ID: mdl-362675
RESUMO
These data presented here demonstrate how careful analysis of mock loop testing can lead to useful measurements for long-term calf experiments. The accuracy of the data rae primarily dependent upon a valid circulatory system analog and good experimental technique. These methods of determining arterial pressure, left atrial pressure and cardiac output have allowed us to obtain recordds of these important parameters for periods as long as 100 days in calves with total artificial heart implantation. These methods have also enabled us to automatically control the artifical heart under conditions of rest, exercise and pharmacologic treatment with the use of only one external pneumatic drive line pressure transducer.
Assuntos
Pressão Sanguínea , Débito Cardíaco , Coração Artificial , Animais , Bovinos
2.
Am J Nurs
; 66(8): 1782-3, 1966 Aug.
Artigo
em Inglês
| MEDLINE
| ID: mdl-5178210