1.
Sci Rep ; 12(1): 16532, 2022 10 03.
Artigo em Inglês | MEDLINE | ID: mdl-36192423

RESUMO

Wearable sensors have seen remarkable recent technological developments, and their role in healthcare is expected to expand. Specifically, monitoring tissue circulation in patients who have undergone reconstructive surgery is critical because blood flow deficiencies must be rescued within hours or the transplant will fail due to thrombosis/haematoma within the artery or vein. We design a wearable, wireless, continuous, multipoint sensor to monitor tissue circulation. The system measures pulse waves, skin colour, and tissue temperature to reproduce physician assessment. Data are analysed in real time for patient risk using an algorithm. This multicentre clinical trial involved 73 patients who underwent transplant surgery and had their tissue circulation monitored until postoperative day 7. Herein, we show that the overall agreement rate between physician and sensor findings is 99.2%. In addition, the patient questionnaire results indicate that the device is easy to wear. The sensor demonstrates non-invasive, real-time, continuous, multi-point, wireless, and reliable monitoring for postoperative care. This wearable system can improve the success rate of reconstructive surgeries.


Assuntos
Dispositivos Eletrônicos Vestíveis , Artérias , Frequência Cardíaca , Humanos , Monitorização Fisiológica , Cuidados Pós-Operatórios
SELEÇÃO DE REFERÊNCIAS