Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros










Intervalo de ano de publicação
1.
Artigo em Chinês | WPRIM (Pacífico Ocidental) | ID: wpr-260092

RESUMO

Current permanent fight ventricular and fight atrial endocardial pacing leads are implanted utilizing a central lumen stylet. Right ventricular apex pacing initiates an abnormal asyno chronous electrical activation pattern, which results in asynchronous ventricular contraction and re-laxation. When pacing from right atrial appendage, the conduction time between two atria will be prolonged, which results in heterogeneity for both depolarization and repolarization. Six patients with Class Ⅰ indication for permanent pacing were implanted with either single chamber or dual chamber pacemaker. The SelectSecure 3830 4-French (Fr) lumenless lead and the SelectSite C304 8.5-Fr steerable catheter-sheath (Medtronic Inc., USA) were used. Pre-selected pacing sites included inter-atrial septum and right ventricular outflow tract, which were defined by ECG and fluoroscopic criteria. All the implanting procedures were successful without complication. Testing results (mean atrial pacing threshold: 0.87 V; mean P wave amplitude: 2.28 mV; mean ventricular pacing threshold:0.53V; mean R wave amplitude: 8.75 mV) were satisfactory. It is concluded that implantation of a 4-Fr lumenless pacing lead by using a streerable catheter-sheath to achieve inter-atrial septum or right ventricular outflow tract pacing is safe and feasible.

SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...