Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 4 de 4
Filtrar
1.
J AOAC Int ; 102(2): 508-524, 2019 Mar 01.
Artigo em Inglês | MEDLINE | ID: mdl-30205867

RESUMO

VereBeef™ Detection Kit, incorporating both multiplex PCR and microarray technologies on a lab-on-chip platform, is intended for qualitative detection and differentiation of Escherichia coli O157:H7, E. coli O26, E. coli O45, E. coli O103, E. coli O111, E. coli O121, E. coli O145, Shiga toxin-producing E. coli (STEC) virulence factors (stx1A, stx2A, eae), and Salmonella species in one test using raw beef trim samples. This product underwent extensive evaluations, including inclusivity-exclusivity, method comparison, robustness, lot-to-lot variability, and stability studies. The inclusivity/exclusivity study demonstrated that VereBeef Detection Kit specifically detects and identifies target analytes without occurrence of false-positive and false-negative detection. In the method comparison study, the performance of the VereBeef Detection Kit was compared with U.S. Department of Agriculture Food Safety and Inspection Service Microbiology Laboratory Guidebook's methods for target organism detection in raw beef trim using E. coli O157:H7 single inoculation and Salmonella and non-O157 STEC dual inoculation. Data demonstrated equivalence in both methods. The robustness study showed that changes in the test parameters do not impact assay performance. Collectively, VereBeef Detection Kit is able to detect target pathogens in raw beef trim with a minimum enrichment time of 8 h for E. coli O157:H7 detection and 10 h for Salmonella and non-O157 STEC detection.


Assuntos
Microbiologia de Alimentos , Carne/microbiologia , Técnicas Analíticas Microfluídicas/normas , Reação em Cadeia da Polimerase Multiplex/normas , Animais , Bovinos , Escherichia coli/classificação , Escherichia coli/isolamento & purificação , Salmonella/classificação , Salmonella/isolamento & purificação
2.
PLoS Negl Trop Dis ; 8(7): e3043, 2014.
Artigo em Inglês | MEDLINE | ID: mdl-25078474

RESUMO

Tropical pathogens often cause febrile illnesses in humans and are responsible for considerable morbidity and mortality. The similarities in clinical symptoms provoked by these pathogens make diagnosis difficult. Thus, early, rapid and accurate diagnosis will be crucial in patient management and in the control of these diseases. In this study, a microfluidic lab-on-chip integrating multiplex molecular amplification and DNA microarray hybridization was developed for simultaneous detection and species differentiation of 26 globally important tropical pathogens. The analytical performance of the lab-on-chip for each pathogen ranged from 102 to 103 DNA or RNA copies. Assay performance was further verified with human whole blood spiked with Plasmodium falciparum and Chikungunya virus that yielded a range of detection from 200 to 4×105 parasites, and from 250 to 4×107 PFU respectively. This lab-on-chip was subsequently assessed and evaluated using 170 retrospective patient specimens in Singapore and Thailand. The lab-on-chip had a detection sensitivity of 83.1% and a specificity of 100% for P. falciparum; a sensitivity of 91.3% and a specificity of 99.3% for P. vivax; a positive 90.0% agreement and a specificity of 100% for Chikungunya virus; and a positive 85.0% agreement and a specificity of 100% for Dengue virus serotype 3 with reference methods conducted on the samples. Results suggested the practicality of an amplification microarray-based approach in a field setting for high-throughput detection and identification of tropical pathogens.


Assuntos
Doenças Transmissíveis/diagnóstico , Dispositivos Lab-On-A-Chip , Técnicas Analíticas Microfluídicas/métodos , Técnicas de Diagnóstico Molecular/métodos , Medicina Tropical/métodos , Humanos , Técnicas Analíticas Microfluídicas/instrumentação , Técnicas de Diagnóstico Molecular/instrumentação , Sensibilidade e Especificidade , Singapura , Tailândia , Medicina Tropical/instrumentação
3.
Arch Virol ; 156(8): 1371-8, 2011 Aug.
Artigo em Inglês | MEDLINE | ID: mdl-21503642

RESUMO

Threatening sporadic outbreaks of avian influenza and the H1N1 pandemic of 2009 highlight the need for rapid and accurate detection and typing of influenza viruses. In this paper, we describe the validation of the VereFlu™ Lab-on-Chip Influenza Assay, which is based on the integration of two technologies: multiplex reverse transcription (RT)-PCR followed by microarray amplicon detection. This assay simultaneously detects five influenza virus subtypes, including the 2009 pandemic influenza A (H1N1), seasonal H1N1, H3N2, H5N1 and influenza B virus. The VereFlu™ assay was clinically validated in Singapore and compared against reference methods of real-time PCR, virus detection by immunofluorescence of cell cultures and sequencing. A sensitivity and specificity of 96.8% and 92.8%, respectively, was demonstrated for pandemic H1N1; 95.7% and 100%, respectively, for seasonal H1N1; 91.2% and 97.6%, respectively, for seasonal H3N2; 95.2% and 100%, respectively, for influenza B. Additional evaluations carried out at the World Health Organization (WHO) Collaborating Centre, Melbourne, Australia, confirmed that the test was able to reliably detect H5N1. This portable, fast time-to-answer (3 hours) device is particularly suited for diagnostic applications of detection, differentiation and identification of human influenza virus subtypes.


Assuntos
Vírus da Influenza A/isolamento & purificação , Vírus da Influenza B/isolamento & purificação , Análise Serial de Proteínas/métodos , Reação em Cadeia da Polimerase Via Transcriptase Reversa/métodos , Humanos , Vírus da Influenza A/classificação , Dispositivos Lab-On-A-Chip , Reprodutibilidade dos Testes , Sensibilidade e Especificidade
4.
BMC Infect Dis ; 6: 40, 2006 Mar 02.
Artigo em Inglês | MEDLINE | ID: mdl-16512903

RESUMO

BACKGROUND: Continuous outbreaks of the highly pathogenic H5N1 avian influenza A in Asia has resulted in an urgent effort to improve current diagnostics to aid containment of the virus and lower the threat of a influenza pandemic. We report here the development of a PCR-based assay that is highly specific for the H5N1 avian influenza A virus. METHODS: A one-step reverse-transcription PCR assay was developed to detect the H5N1 avian influenza A virus. The specificity of the assay was shown by testing sub-types of influenza A virus and other viral and bacterial pathogens; and on field samples. RESULTS: Validation on 145 field specimens from Vietnam and Malaysia showed that the assay was specific without cross reactivity to a number of other infuenza strains as well as human respiratory related pathogens. Detection was 100% from allantoic fluid in H5N1 positive samples, suggesting it to be a reliable sampling source for accurate detection. CONCLUSION: The assay developed from this study indicates that the primers are specific for the H5N1 influenza virus. As shown by the field tested results, this assay would be highly useful as a diagnostic tool to help identify and control influenza epidemics.


Assuntos
Virus da Influenza A Subtipo H5N1/genética , Virus da Influenza A Subtipo H5N1/isolamento & purificação , Reação em Cadeia da Polimerase Via Transcriptase Reversa/métodos , Reações Cruzadas , Humanos , Reprodutibilidade dos Testes , Sensibilidade e Especificidade
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...