1.
IEEE Trans Biomed Circuits Syst ; 15(2): 303-313, 2021 04.
Artigo em Inglês | MEDLINE | ID: mdl-33760740

RESUMO

In the intact brain, neural activity can be recorded using sensing electrodes and manipulated using light stimulation. Silicon probes with integrated electrodes and µLEDs enable the detection and control of neural activity using a single implanted device. Miniaturized solutions for recordings from small freely moving animals are commercially available, but stimulation is driven by large, stationary current sources. We designed and fabricated a current source chip and integrated it into a headstage PCB that weighs 1.37 g. The proposed system provides 10-bit resolution current control for 32 channels, driving µLEDs with up to 4.6 V and sourcing up to 0.9 mA at a refresh rate of 5 kHz per channel. When calibrated against a µLED probe, the system allows linear control of light output power, up to 10 µW per µLED. To demonstrate the capabilities of the system, synthetic sequences of neural spiking activity were produced by driving multiple µLEDs implanted in the hippocampal CA1 area of a freely moving mouse. The high spatial, temporal, and amplitude resolution of the system provides a rich variety of stimulation patterns. Combined with commercially available sampling headstages, the system provides an easy to use back-end, fully utilizing the bi-directional potential of integrated opto-electronic arrays.


Assuntos
Próteses e Implantes , Silício , Animais , Eletrodos , Hipocampo , Camundongos
SELEÇÃO DE REFERÊNCIAS