1.
Opt Express ; 29(3): 3406-3414, 2021 Feb 01.
Artigo em Inglês | MEDLINE | ID: mdl-33770939

RESUMO

For low-order adaptive optics systems, a method that is able to correct for system aberrations in the final focal plane is presented. The paper presents a novel figure of merit, corresponding to the integrated non-normalized tip-tilt-free optical transfer function. The inherent singular value decomposition modal content of the interaction matrix is used to optimize this figure of merit. The method has proven to be stable and robust, providing a simple mean to facilitate diffraction limited imaging in an experimental setup for ophthalmic applications.

2.
Acta Ophthalmol ; 92(6): 557-62, 2014 Sep.
Artigo em Inglês | MEDLINE | ID: mdl-24373288

RESUMO

PURPOSE: The purpose of our study was to establish a practical and quick test for assessing reading performance and to statistically analyse interchart and test-retest reliability of a new standardized Swedish reading chart system consisting of three charts constructed according to the principles available in the literature. METHODS: Twenty-four subjects with healthy eyes, mean age 65 ± 10 years, were tested binocularly and the reading performance evaluated as reading acuity, critical print size and maximum reading speed. The test charts all consist of 12 short text sentences with a print size ranging from 0.9 to -0.2 logMAR in approximate steps of 0.1 logMAR. Two testing sessions, in two different groups (C1 and C2), were under strict control of luminance and lighting environment. Reading performance tests with chart T1, T2 and T3 were used for evaluation of interchart reliability and test data from a second session 1 month or more apart for the test-retest analysis. RESULTS: The testing of reading performance in adult observers with short sentences of continuous text was quick and practical. The agreement between the tests obtained with the three different test charts was high both within the same test session and at retest. CONCLUSION: This new Swedish variant of a standardized reading system based on short sentences and logarithmic progression of print size provides reliable measurements of reading performance and preliminary norms in an age group around 65 years. The reading test with three independent reading charts can be useful for clinical studies of reading ability before and after treatment.


Assuntos
Leitura , Testes Visuais/normas , Visão Binocular/fisiologia , Acuidade Visual/fisiologia , Idoso , Idoso de 80 Anos ou mais , Feminino , Voluntários Saudáveis , Humanos , Idioma , Masculino , Pessoa de Meia-Idade , Reprodutibilidade dos Testes , Suécia , Testes Visuais/instrumentação
3.
Invest Ophthalmol Vis Sci ; 52(5): 2649-55, 2011 Apr 22.
Artigo em Inglês | MEDLINE | ID: mdl-21228372

RESUMO

PURPOSE: To demonstrate noninvasive imaging of human foveal capillary networks with a high-resolution, wide-field, dual-conjugate adaptive optics (DCAO) imaging instrument. METHODS: The foveal capillary networks of five healthy subjects with no previous history of ocular or neurologic disease or surgery were imaged with a novel high-resolution, wide-field DCAO instrument. The foveal avascular zone (FAZ) in each image was defined using a manual procedure. An automated algorithm based on publicly available and custom-written software was used to identify vessels and extract morphologic FAZ and vessel parameters. Capillary densities were calculated in two annular regions of interest (ROIs) outside the FAZ (500 µm and 750 µm outer radius from the foveal center) and in the superior, inferior, temporal, and nasal quadrants within the two ROIs. RESULTS: Mean FAZ area was 0.302 ± 0.100 mm(2), and mean capillary density (length/area) in the inner ROI was 38.0 ± 4.0 mm(-1) and 36.4 ± 4.0 mm(-1) in the outer ROI. The difference in ROI capillary density was not significant. There was no significant difference in quadrant capillary density within the two ROIs or between quadrants irrespective of ROI. CONCLUSIONS: The authors have demonstrated a technique for noninvasive imaging and semiautomated detection and analysis of foveal capillaries. In comparison with other studies, their method yielded lower capillary densities than histology but similar results to the current clinical gold standard, fluorescein angiography. The increased field of view of the DCAO instrument opens up new possibilities for high-resolution noninvasive clinical imaging of foveal capillaries.


Assuntos
Diagnóstico por Imagem/instrumentação , Técnicas de Diagnóstico Oftalmológico , Fóvea Central/irrigação sanguínea , Vasos Retinianos/anatomia & histologia , Adulto , Capilares/anatomia & histologia , Humanos , Pessoa de Meia-Idade , Óptica e Fotônica/instrumentação
4.
Opt Express ; 17(6): 4454-67, 2009 Mar 16.
Artigo em Inglês | MEDLINE | ID: mdl-19293873

RESUMO

We present analysis and preliminary laboratory testing of a real-time dual-conjugate adaptive optics (DCAO) instrument for ophthalmology that will enable wide-field high resolution imaging of the retina in vivo. The setup comprises five retinal guide stars (GS) and two deformable mirrors (DM), one conjugate to the pupil and one conjugate to a plane close to the retina. The DCAO instrument has a closed-loop wavefront sensing wavelength of 834 nm and an imaging wavelength of 575 nm. It incorporates an array of collimator lenses to spatially filter the light from all guide stars using one adjustable iris, and images the Hartmann patterns of multiple reference sources on a single detector. Zemax simulations were performed at 834 nm and 575 nm with the Navarro 99 and the Liou- Brennan eye models. Two correction alternatives were evaluated; conventional single conjugate AO (SCAO, using one GS and a pupil DM) and DCAO (using multiple GS and two DM). Zemax simulations at 575 nm based on the Navarro 99 eye model show that the diameter of the corrected field of view for diffraction-limited imaging (Strehl >or= 0.8) increases from 1.5 deg with SCAO to 6.5 deg using DCAO. The increase for the less stringent condition of a wavefront error of 1 rad or less (Strehl >or= 0.37) is from 3 deg with SCAO to approximately 7.4 deg using DCAO. Corresponding results for the Liou-Brennan eye model are 3.1 deg (SCAO) and 8.2 deg (DCAO) for Strehl >or= 0.8, and 4.8 deg (SCAO) and 9.6 deg (DCAO) for Strehl >or= 0.37. Potential gain in corrected field of view with DCAO is confirmed both by laboratory experiments on a model eye and by preliminary in vivo imaging of a human eye.


Assuntos
Imageamento Tridimensional/métodos , Oftalmologia/instrumentação , Oftalmologia/métodos , Óptica e Fotônica/instrumentação , Retina/anatomia & histologia , Simulação por Computador , Fóvea Central/anatomia & histologia , Fundo de Olho , Humanos , Análise Espectral
5.
J Opt Soc Am A Opt Image Sci Vis ; 19(1): 152-7, 2002 Jan.
Artigo em Inglês | MEDLINE | ID: mdl-11778718

RESUMO

The purpose of this work is to investigate the light-scattering properties of excised eye lenses with intact lens capsules--more specifically, to compare light scattering with light transmission at different wavelengths in aging and cataractous lenses. A lens is positioned at its focal-length distance from an optical fiber end and collimates the light from one of five laser lines (458-633 nm). By use of an integrating sphere with an extra circular port, the collimated directly transmitted light can be separated from the scattered light. For lenses with low light-scattering levels, integrated scattering showed a dependence on wavelength, but when light scattering increased the wavelength difference tended to level out. Despite the higher percentage of lens light scattering at lower wavelengths, when calculated as an "effective light scattering" (compensated for light transmission), more scattered light actually falls toward the retina at longer wavelengths.