Curr Issues Mol Biol ; 45(12): 9354-9367, 2023 Nov 22.
Artigo em Inglês | MEDLINE | ID: mdl-38132432

RESUMO

In neonatal screening, amino acids have a significant diagnostic role. Determination of their values may identify abnormal conditions. Early diagnosis and continuous monitoring of amino acid disorders results in a better disease outcome. An easy and simple LC-MS/MS method was developed for the quantitation of underivatized amino acids. Amino acids were separated using a normal-phase HPLC column having a totally porous silica stationary phase and using classical reversed-phase eluents. Mass spectrometry in multiple reaction monitoring mode was used for the analysis, providing high selectivity and sensitivity. A standard addition calibration model was applied for quantitation using only one isotope-labeled internal standard for all amino acids. Five calibration points were used for quantitation, and the method was successfully validated. The slopes of the calibration curves of the individual amino acids in parallel measurements were found to be similar. Since the measured slopes were reproducible, one serum sample could represent every series of serum samples of a given day. The method was tested on human serum samples and adequate results were obtained. This new method can be easily applied in clinical laboratories.