1.
Radiat Prot Dosimetry ; 131(3): 374-8, 2008.
Artigo em Inglês | MEDLINE | ID: mdl-18586862

RESUMO

The objective of this study is to evaluate the precision of dose-calculation computer codes used in our laboratory (PCXMC and PREPARE) for organ dose evaluation. Measurements of entrance and organ dose were performed using ionisation chamber and thermoluminescence dosimetry. To obtain a mean dose of organ, we have used the Rando-Alderson phantom. The results showed that computed and measured doses correlate well (within 28%) in 60% of the samples. The percentage shows that the computed doses correlate with the experimental doses rather well for PCXMC software than PREPARE. Although the two programs are based on the Monte-Carlo method, their calculations differ. PCXMC carries out a simulation of the trajectory of the photon, whereas PREPARE provides interpolated values. Our experimental results are close to the values given by the PCXMC, a program which takes into account the weight, the height of the patient and field dimensions.


Assuntos
Simulação por Computador , Método de Monte Carlo , Imagens de Fantasmas , Radiologia/métodos , Radiometria/instrumentação , Radiometria/métodos , Algoritmos , Humanos , Pulmão/diagnóstico por imagem , Radiografia , Tórax/efeitos da radiação