1.
Blood Press Monit ; 24(3): 151-154, 2019 Jun.
Artigo em Inglês | MEDLINE | ID: mdl-30807305

RESUMO

The objective of this study was to determine the accuracy of the Hingmed WBP-02A device for ambulatory blood pressure (BP) monitoring according to the International Protocol of the European Society of Hypertension (ESH-IP). Device evaluation was carried out in 33 patients. The mean age was 69.8±13.8 years, the mean systolic BP was 137.6±27.0 mmHg (range: 96-172), the mean diastolic BP was 86.7±20.1 mmHg (range: 48-124) and the mean arm circumference was 30.1±6.2 cm (range: 20-43). The protocol requirements were followed precisely. Four different cuffs, small, standard, large, and extra-large were used. The device comfortably passed all ESH-IP requirements, as 84 out of 99 device-observer systolic BP differences and 97 of 99 diastolic BP differences were within ±5 mmHg thereby fulfilling the standards of the protocol. Device-observer BP differences were unrelated to arm circumference. These data show that the Hingmed WBP-02A device for ambulatory BP monitoring met the requirements for validation by the ESH-IP and can be recommended for clinical use in the adult population.


Assuntos
Monitorização Ambulatorial da Pressão Arterial/instrumentação , Monitores de Pressão Arterial/normas , Idoso , Idoso de 80 Anos ou mais , Pressão Sanguínea , Diástole , Feminino , Humanos , Hipertensão/diagnóstico , Masculino , Pessoa de Meia-Idade , Sístole
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...