Biomed Chromatogr ; 34(5): e4812, 2020 May.
Artigo em Inglês | MEDLINE | ID: mdl-32068899

RESUMO

In this research, we developed and validated a liquid chromatography coupled to mass spectrometry (LC-QToF-MS) method for simultaneous quantification of the anti-tuberculosis drugs ethambutol, isoniazid, pyrazinamide and rifampicin in human plasma. Plasma samples spiked with cimetidine (internal standard) were extracted using protein precipitation with acetonitrile containing 1% formic acid. Separation was performed using a C18 column under flow gradient conditions with water and acetonitrile, both containing 5 mm ammonium formate and 0.1% formic acid. The method was validated according to the ANVISA and US Food and Drug Administration guidelines for bioanalytical method validation. The calibration curve was linear over a concentration range of 0.2-5 µg ml-1 for ethambutol, 0.2-7.5 µg ml-1 for isoniazid, 1-40 µg ml-1 for pyrazinamide and 0.25-2 µg ml-1 for rifampicin, all with adequate precision and accuracy. The method was reproducible, selective and free of carryover and matrix effects. The validated LC-QToF-MS method was successfully applied to real samples and shown to be applicable to future therapeutic and pharmacokinetic monitoring studies.


Assuntos
Antituberculosos/sangue , Cromatografia Líquida de Alta Pressão/métodos , Etambutol/sangue , Isoniazida/sangue , Espectrometria de Massas/métodos , Pirazinamida/sangue , Rifampina/sangue , Humanos , Plasma/química