Eur Radiol ; 30(2): 934-942, 2020 Feb.
Artigo em Inglês | MEDLINE | ID: mdl-31471752

RESUMO

OBJECTIVES: To evaluate the accuracy and clinical integrability of a comprehensive simulation tool to plan and predict radiofrequency ablation (RFA) zones in liver tumors. METHODS: Forty-five patients with 51 malignant hepatic lesions of different origins were included in a prospective multicenter trial. Prior to CT-guided RFA, all patients underwent multiphase CT which included acquisitions for the assessment of liver perfusion. These data were used to generate a 3D model of the liver. The intra-procedural position of the RFA probe was determined by CT and semi-automatically registered to the 3D model. Size and shape of the simulated ablation zones were compared with those of the thermal ablation zones segmented in contrast-enhanced CT images 1 month after RFA; procedure time was compared with a historical control group. RESULTS: Simulated and segmented ablation zone volumes showed a significant correlation (ρ = 0.59, p < 0.0001) and no significant bias (Wilcoxon's Z = 0.68, p = 0.25). Representative measures of ablation zone comparison were as follows: average surface deviation (absolute average error, AAE) with 3.4 ± 1.7 mm, Dice similarity coefficient 0.62 ± 0.14, sensitivity 0.70 ± 0.21, and positive predictive value 0.66 ± 0. There was a moderate positive correlation between AAE and duration of the ablation (∆t; r = 0.37, p = 0.008). After adjustments for inter-individual differences in ∆t, liver perfusion, and prior transarterial chemoembolization procedures, ∆t was an independent predictor of AAE (ß = 0.03 mm/min, p = 0.01). Compared with a historical control group, the simulation added 3.5 ± 1.9 min to the procedure. CONCLUSION: The validated simulation tool showed acceptable speed and accuracy in predicting the size and shape of hepatic RFA ablation zones. Further randomized controlled trials are needed to evaluate to what extent this tool might improve patient outcomes. KEY POINTS: • More reliable, patient-specific intra-procedural estimation of the induced RFA ablation zones in the liver may lead to better planning of the safety margins around tumors. • Dedicated real-time simulation software to predict RFA-induced ablation zones in patients with liver malignancies has shown acceptable agreement with the follow-up results in a first prospective multicenter trial suggesting a randomized controlled clinical trial to evaluate potential outcome benefit for patients.


Assuntos
Carcinoma Hepatocelular/cirurgia , Ablação por Cateter/métodos , Neoplasias Hepáticas/cirurgia , Adolescente , Adulto , Idoso , Carcinoma Hepatocelular/patologia , Quimioembolização Terapêutica/métodos , Simulação por Computador , Feminino , Humanos , Fígado/patologia , Fígado/cirurgia , Neoplasias Hepáticas/patologia , Masculino , Margens de Excisão , Pessoa de Meia-Idade , Planejamento de Assistência ao Paciente , Estudos Prospectivos , Tomografia Computadorizada por Raios X , Adulto Jovem
2.
Sci Rep ; 8(1): 787, 2018 01 15.
Artigo em Inglês | MEDLINE | ID: mdl-29335429

RESUMO

The RFA Guardian is a comprehensive application for high-performance patient-specific simulation of radiofrequency ablation of liver tumors. We address a wide range of usage scenarios. These include pre-interventional planning, sampling of the parameter space for uncertainty estimation, treatment evaluation and, in the worst case, failure analysis. The RFA Guardian is the first of its kind that exhibits sufficient performance for simulating treatment outcomes during the intervention. We achieve this by combining a large number of high-performance image processing, biomechanical simulation and visualization techniques into a generalized technical workflow. Further, we wrap the feature set into a single, integrated application, which exploits all available resources of standard consumer hardware, including massively parallel computing on graphics processing units. This allows us to predict or reproduce treatment outcomes on a single personal computer with high computational performance and high accuracy. The resulting low demand for infrastructure enables easy and cost-efficient integration into the clinical routine. We present a number of evaluation cases from the clinical practice where users performed the whole technical workflow from patient-specific modeling to final validation and highlight the opportunities arising from our fast, accurate prediction techniques.


Assuntos
Ablação por Cateter , Neoplasias Hepáticas/cirurgia , Cirurgia Assistida por Computador/métodos , Humanos , Processamento de Imagem Assistida por Computador , Neoplasias Hepáticas/patologia , Tomografia Computadorizada por Raios X , Resultado do Tratamento
3.
Contemp Clin Trials Commun ; 8: 25-32, 2017 Dec.
Artigo em Inglês | MEDLINE | ID: mdl-29696193

RESUMO

INTRODUCTION: Radio-frequency ablation (RFA) is a promising minimal-invasive treatment option for early liver cancer, however monitoring or predicting the size of the resulting tissue necrosis during the RFA-procedure is a challenging task, potentially resulting in a significant rate of under- or over treatments. Currently there is no reliable lesion size prediction method commercially available. OBJECTIVES: ClinicIMPPACT is designed as multicenter-, prospective-, non-randomized clinical trial to evaluate the accuracy and efficiency of innovative planning and simulation software. 60 patients with early liver cancer will be included at four European clinical institutions and treated with the same RFA system. The preinterventional imaging datasets will be used for computational planning of the RFA treatment. All ablations will be simulated simultaneously to the actual RFA procedure, using the software environment developed in this project. The primary outcome measure is the comparison of the simulated ablation zones with the true lesions shown in follow-up imaging after one month, to assess accuracy of the lesion prediction. DISCUSSION: This unique multicenter clinical trial aims at the clinical integration of a dedicated software solution to accurately predict lesion size and shape after radiofrequency ablation of liver tumors. Accelerated and optimized workflow integration, and real-time intraoperative image processing, as well as inclusion of patient specific information, e.g. organ perfusion and registration of the real RFA needle position might make the introduced software a powerful tool for interventional radiologists to optimize patient outcomes.

4.
Int J Comput Assist Radiol Surg ; 12(1): 59-68, 2017 Jan.
Artigo em Inglês | MEDLINE | ID: mdl-27538836

RESUMO

PURPOSE: Radiofrequency ablation (RFA) is one of the most popular and well-standardized minimally invasive cancer treatments (MICT) for liver tumours, employed where surgical resection has been contraindicated. Less-experienced interventional radiologists (IRs) require an appropriate planning tool for the treatment to help avoid incomplete treatment and so reduce the tumour recurrence risk. Although a few tools are available to predict the ablation lesion geometry, the process is computationally expensive. Also, in our implementation, a few patient-specific parameters are used to improve the accuracy of the lesion prediction. METHODS: Advanced heterogeneous computing using personal computers, incorporating the graphics processing unit (GPU) and the central processing unit (CPU), is proposed to predict the ablation lesion geometry. The most recent GPU technology is used to accelerate the finite element approximation of Penne's bioheat equation and a three state cell model. Patient-specific input parameters are used in the bioheat model to improve accuracy of the predicted lesion. RESULTS: A fast GPU-based RFA solver is developed to predict the lesion by doing most of the computational tasks in the GPU, while reserving the CPU for concurrent tasks such as lesion extraction based on the heat deposition at each finite element node. The solver takes less than 3 min for a treatment duration of 26 min. When the model receives patient-specific input parameters, the deviation between real and predicted lesion is below 3 mm. CONCLUSION: A multi-centre retrospective study indicates that the fast RFA solver is capable of providing the IR with the predicted lesion in the short time period before the intervention begins when the patient has been clinically prepared for the treatment.


Assuntos
Carcinoma Hepatocelular/cirurgia , Ablação por Cateter/métodos , Gráficos por Computador , Neoplasias Hepáticas/cirurgia , Carcinoma Hepatocelular/diagnóstico por imagem , Simulação por Computador , Humanos , Neoplasias Hepáticas/diagnóstico por imagem , Procedimentos Cirúrgicos Minimamente Invasivos , Modelos Teóricos , Imagem de Perfusão , Estudos Retrospectivos , Cirurgia Assistida por Computador , Tomografia Computadorizada por Raios X
5.
Sci Rep ; 5: 15373, 2015 Oct 20.
Artigo em Inglês | MEDLINE | ID: mdl-26482818

RESUMO

Percutaneous radiofrequency ablation (RFA) is a minimally invasive technique that destroys cancer cells by heat. The heat results from focusing energy in the radiofrequency spectrum through a needle. Amongst others, this can enable the treatment of patients who are not eligible for an open surgery. However, the possibility of recurrent liver cancer due to incomplete ablation of the tumor makes post-interventional monitoring via regular follow-up scans mandatory. These scans have to be carefully inspected for any conspicuousness. Within this study, the RF ablation zones from twelve post-interventional CT acquisitions have been segmented semi-automatically to support the visual inspection. An interactive, graph-based contouring approach, which prefers spherically shaped regions, has been applied. For the quantitative and qualitative analysis of the algorithm's results, manual slice-by-slice segmentations produced by clinical experts have been used as the gold standard (which have also been compared among each other). As evaluation metric for the statistical validation, the Dice Similarity Coefficient (DSC) has been calculated. The results show that the proposed tool provides lesion segmentation with sufficient accuracy much faster than manual segmentation. The visual feedback and interactivity make the proposed tool well suitable for the clinical workflow.


Assuntos
Carcinoma Hepatocelular/diagnóstico , Carcinoma Hepatocelular/terapia , Ablação por Cateter , Neoplasias Hepáticas/diagnóstico , Neoplasias Hepáticas/terapia , Tomografia Computadorizada por Raios X , Algoritmos , Ablação por Cateter/métodos , Humanos , Aumento da Imagem , Estudos Retrospectivos , Máquina de Vetores de Suporte , Tomografia Computadorizada por Raios X/métodos , Resultado do Tratamento
6.
Artigo em Inglês | MEDLINE | ID: mdl-26736783

RESUMO

In this contribution, we present a semi-automatic segmentation algorithm for radiofrequency ablation (RFA) zones via optimal s-t-cuts. Our interactive graph-based approach builds upon a polyhedron to construct the graph and was specifically designed for computed tomography (CT) acquisitions from patients that had RFA treatments of Hepatocellular Carcinomas (HCC). For evaluation, we used twelve post-interventional CT datasets from the clinical routine and as evaluation metric we utilized the Dice Similarity Coefficient (DSC), which is commonly accepted for judging computer aided medical segmentation tasks. Compared with pure manual slice-by-slice expert segmentations from interventional radiologists, we were able to achieve a DSC of about eighty percent, which is sufficient for our clinical needs. Moreover, our approach was able to handle images containing (DSC=75.9%) and not containing (78.1%) the RFA needles still in place. Additionally, we found no statistically significant difference (p<;0.423) between the segmentation results of the subgroups for a Mann-Whitney test. Finally, to the best of our knowledge, this is the first time a segmentation approach for CT scans including the RFA needles is reported and we show why another state-of-the-art segmentation method fails for these cases. Intraoperative scans including an RFA probe are very critical in the clinical practice and need a very careful segmentation and inspection to avoid under-treatment, which may result in tumor recurrence (up to 40%). If the decision can be made during the intervention, an additional ablation can be performed without removing the entire needle. This decreases the patient stress and associated risks and costs of a separate intervention at a later date. Ultimately, the segmented ablation zone containing the RFA needle can be used for a precise ablation simulation as the real needle position is known.


Assuntos
Técnicas de Ablação/instrumentação , Carcinoma Hepatocelular/diagnóstico por imagem , Carcinoma Hepatocelular/cirurgia , Processamento de Imagem Assistida por Computador/métodos , Neoplasias Hepáticas/diagnóstico por imagem , Neoplasias Hepáticas/cirurgia , Agulhas , Ondas de Rádio , Algoritmos , Humanos , Recidiva , Tomografia Computadorizada por Raios X