RESUMO
Conventional approaches for MR elastography (MRE) using surface drivers have difficulty achieving sufficient shear wave propagation in the prostate gland due to attenuation. In this study we evaluate the feasibility of generating shear wave propagation in the prostate gland using a transurethral device. A novel transurethral actuator design is proposed, and the performance of this device was evaluated in gelatin phantoms and in a canine prostate gland. All MRI was performed on a 1.5T MR imager using a conventional gradient-echo MRE sequence. A piezoceramic actuator was used to vibrate the transurethral device along its length. Shear wave propagation was measured transverse and parallel to the rod at frequencies between 100 and 250 Hz in phantoms and in the prostate gland. The shear wave propagation was cylindrical, and uniform along the entire length of the rod in the gel experiments. The feasibility of transurethral MRE was demonstrated in vivo in a canine model, and shear wave propagation was observed in the prostate gland as well as along the rod. These experiments demonstrate the technical feasibility of transurethral MRE in vivo. Further development of this technique is warranted.
Assuntos
Técnicas de Imagem por Elasticidade/instrumentação , Magnetismo/instrumentação , Próstata/anatomia & histologia , Próstata/fisiologia , Transdutores , Animais , Cães , Desenho de Equipamento , Análise de Falha de Equipamento , Humanos , Masculino , Imagens de Fantasmas , Reprodutibilidade dos Testes , Sensibilidade e Especificidade , Uretra/anatomia & histologia , Uretra/fisiologiaRESUMO
The feasibility of using MR thermometry for temperature feedback to control a transurethral ultrasound heating applicator with planar transducers was investigated. The sensitivity of a temperature-based feedback algorithm to spatial (control point area, slice thickness, angular alignment) and non-spatial (imaging time, temperature uncertainty) parameters was evaluated through numerical simulations. The angular alignment of the control point with the ultrasound beam was an important parameter affecting the average spatial error in heat delivery. The other spatial parameters were less influential, thus providing an opportunity to reduce spatial resolution for increased SNR in the MR imaging. The update time was the most important non-spatial parameter determining the performance of the control algorithm. Combined non-spatial and spatial parameters achieved acceptable performance with a voxel size of 3 mm x 3 mm, a 10 mm slice thickness and a 5 s update time. Temperature uncertainty of up to 2 degrees C had little effect on the performance of the control algorithm but did reduce the average error slightly due to a systematic, noise-induced overestimation of the boundary temperature. These simulations imply that MR thermometry performed on clinical 1.5 T imaging systems is of sufficient quality for use as thermal feedback for conformal prostate thermal therapy with transurethral ultrasound heating applicators incorporating planar transducers.