1.
Rev Sci Instrum ; 81(10): 10D319, 2010 Oct.
Artigo em Inglês | MEDLINE | ID: mdl-21033845

RESUMO

The first 3 of 18 neutron time-of-flight (nTOF) channels have been installed at the National Ignition Facility (NIF). The role of these detectors includes yield, temperature, and bang time measurements. This article focuses on nTOF data analysis and quality of results obtained for the first set of experiments to use all 192 NIF beams. Targets produced up to 2×10(10) 2.45 MeV neutrons for initial testing of the nTOF detectors. Differences in neutron scattering at the OMEGA laser facility where the detectors were calibrated and at NIF result in different response functions at the two facilities. Monte Carlo modeling shows this difference. The nTOF performance on these early experiments indicates that the nTOF system with its full complement of detectors should perform well in future measurements of yield, temperature, and bang time.