1.
Cureus ; 14(11): e31209, 2022 Nov.
Artigo em Inglês | MEDLINE | ID: mdl-36382325

RESUMO

Premature ventricular contractions (PVCs) that comprise more than 15% of total heartbeats can induce cardiomyopathy in patients with systolic dysfunction, and cardiac ablation is frequently used to reduce PVCs in this patient group. However, cardiac ablation is not entirely without hazards. We report a noninvasive method that dramatically reduced PVCs in a cardiac pacemaker patient from 31% to 3% in seven days by increasing the lower limit pacing rate from 50 beats per minute (bpm) to 60 bpm. Not only were our patient's PVCs reduced by the initial pacing elevation, but PVC levels were maintained below 5% even after the pacemaker's lower limit was returned to its original value of 50 bpm. This irreversible suppression of PVC activity following a three-month pacing elevation is a novel result that might be caused by ventricular remodeling of the original ectopic focus.