Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
Atmosphere (Basel) ; 10(5): 287, 2019.
Artigo em Inglês | MEDLINE | ID: mdl-32704395

RESUMO

To evaluate the feasibility of the Sunset semicontinuous organic and elemental carbon (OC/EC) monitor, the U.S. Environmental Protection Agency (EPA) sponsored the deployment of this monitor at Chemical Speciation Network (CSN) sites with OC and EC measurements via quartz fiber filter collection in Chicago, Illinois; Houston, Texas; Las Vegas, Nevada; St. Louis, Missouri; Rubidoux, California; and Washington, D.C. Houston, St. Louis, and Washington also had collocated Aethalometer black carbon (BC) measurements. Sunset OC generally compared well with the CSN OC (r2 = 0.73 across five sites); the Sunset/CSN OC ratio was, on average, 1.06, with a range among sites of 0.96 to 1.12. Sunset thermal EC and CSN EC did not compare as well, with an overall r2 of 0.22, in part because 26% of the hourly Sunset EC measurements were below the detection limit. Sunset optical EC had a much better correlation to CSN EC (r2 = 0.67 across all sites), with an average Sunset/CSN ratio of 0.90 (range of 0.7 to 1.08). There was also a high correlation of Sunset optical EC with Aethalometer BC (r2 = 0.77 across all sites), though with a larger bias (average Sunset/Aethalometer ratio of 0.56). When the Sunset instrument was working well, OC and OptEC data were comparable to CSN OC and EC.

SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...