1.
Neurodiagn J ; 59(3): 152-162, 2019.
Artigo em Inglês | MEDLINE | ID: mdl-31251123

RESUMO

Triggered electromyography (t-EMG) is a common technique used during spinal instrumentation in surgery for adolescent idiopathic scoliosis. This study tests the validity of t-EMG values obtained with a standard ball-tipped probe after completion of screw placement versus t-EMG values obtained during screw insertion with a powered screwdriver. t-EMG values were collected for screws spanning T7-L5 using both a standard probe and a powered screwdriver. A power analysis determined that a sample size of 300 screws would provide enough precision to estimate limits of agreement within ±2 mA. A monopolar constant current stimulation technique (0.2 ms duration and 3.11 Hz stimulation rate) was used at each level. EMG was acquired with placement of bipolar pairs of subdermal needle electrodes. A Bland-Altman plot was used to assess agreement between threshold readings from the two techniques. Twenty-nine patients were enrolled in this study with 305 screws. t-EMG values measured using a powered screwdriver were on average 1 mA lower than values from a standard probe. When readings less than or equal to 20 mA were considered, the limits of agreement were approximately 4 and 7 mA overall. In total, 28/305 (9%) screws were removed and reinserted, 9/305 (3%) screws were redirected, and 3/305 (1%) screws were aborted based on t-EMG readings. Despite a small overall difference in t-EMG value between the standard probe and screwdriver, there was still large variability in agreement between the two techniques. t-EMG values obtained with a powered screwdriver during screw insertion are not interchangeable with values measured by a probe.


Assuntos
Eletromiografia/métodos , Monitorização Neurofisiológica Intraoperatória/métodos , Parafusos Pediculares , Escoliose/cirurgia , Instrumentos Cirúrgicos , Adolescente , Criança , Estimulação Elétrica , Potenciais Somatossensoriais Evocados , Feminino , Humanos , Masculino , Estudos Prospectivos , Fusão Vertebral/métodos , Resultado do Tratamento , Adulto Jovem
2.
Clin Ophthalmol ; 5: 1733-5, 2011.
Artigo em Inglês | MEDLINE | ID: mdl-22205833

RESUMO

PURPOSE: To report a case of bilateral iritis and transillumination defects after laser hair removal of the eyebrows with an alexandrite laser. METHODS: A 41-year-old male presented with bilateral eye pain and mild photophobia 2 days after receiving alexandrite (755 nm) laser epilation of both eyebrows. Examination showed visual acuity of 20/20 in both eyes, 2+ conjunctival injection in both eyes, 1+ cells in the anterior chamber of right eye and trace cells in left eye, poor right pupil dilation, and left pupil without movement. Intraocular pressure and fundus examination were normal. He was diagnosed with iritis and iris atrophy, associated with laser epilation. Topical steroids and cycloplegic drops were prescribed for 1 month. RESULTS: After 1 month of treatment, transillumination defects remained in both eyes, but greater in right. In dim light, the right pupil was 4 mm and oval and the left pupil was 6 mm and round. Visual acuity remained 20/20 in both eyes. CONCLUSION: Laser hair removal of the eyebrows can lead to permanent ocular damage even with eye protection, and should be avoided.

SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...