Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros











Base de dados
Intervalo de ano de publicação
1.
Artif Organs ; 27(1): 108-13, 2003 Jan.
Artigo em Inglês | MEDLINE | ID: mdl-12534722

RESUMO

The VentrAssist implantable rotary blood pump (IRBP) is an implantable centrifugal blood pump with a hydrodynamically suspended impeller; optimal efficiency requires small running clearances (70-300 microm). The effect of running clearance and polish on hemolysis was evaluated in vitro. Three different human blood suspensions were compared: phosphate buffered saline (PBS), plasma volume expander (Hemaccel), and whole blood. The test conditions were: blood hematocrit 30%, flow rate 5 L/min, pressure across pump 100 mm Hg, 6 h flow period, and 37 degrees C. Normalized Index of Hemolysis (NIH) for the Biomedicus BP-80, used as a control, was: 0.0040 +/- 0.0023 (n = 9; x +/- SD) and 0.00014 +/- 0.00009 (n = 5) for pooled blood suspensions in PBS and Hemaccel respectively, and 0.00053 +/- 0.0002 (n = 3) in whole blood. Hemolysis was reduced by improved surface finish and unaffected by running clearance. NIH for the VentrAssist IRBP with 0.2 microm Ra surface finish was 0.000167 +/- 0.00007 (n = 4) g/100 L in whole human blood, demonstrating minimal hemolysis.


Assuntos
Centrifugação , Eritrócitos/fisiologia , Coração Auxiliar , Hemólise/fisiologia , Meios de Cultura/farmacologia , Eritrócitos/efeitos dos fármacos , Gelatina/farmacologia , Hemodinâmica/efeitos dos fármacos , Hemodinâmica/fisiologia , Hemólise/efeitos dos fármacos , Humanos , Técnicas In Vitro , Polímeros/farmacologia , Desenho de Prótese , Suspensões/farmacologia
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA