Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
J Cardiovasc Electrophysiol ; 27(10): 1199-1205, 2016 10.
Artigo em Inglês | MEDLINE | ID: mdl-27434039

RESUMO

INTRODUCTION: The safety and efficacy of a novel family of quadripolar left ventricular (LV) pacing leads designed to pace from nonapical regions of the LV with low pacing capture thresholds was studied in patients undergoing implantation of a cardiac resynchronization therapy defibrillator (CRT-D). METHODS AND RESULTS: Patients receiving a CRT-D were implanted with 1 of 3 ACUITY X4 leads (Spiral Long, Spiral Short, or Straight), designed to address coronary venous anatomical variability. Electrical performance and LV lead related complications were evaluated 3 and 6 months post implantation, respectively. 764 patients (68 ± 11 years, 66% male) were enrolled; 738 (97%) successfully implanted with an ACUITY X4 lead (Spiral L, n = 239, 31%; Spiral S, n = 281, 37%; Straight, n = 218, 29%). A targeted threshold ≤2.5 V was achieved in 644 (94%) patients. The median threshold from the best proximal electrode was lower than the tip electrode (0.9 V [IQR 0.7, 1.3] vs. 1.3 V [IQR 0.7, 2.5], p< 0.001) on Spiral leads. Irrespective of lead implanted, one of the proximal electrodes was the programmed cathode in most patients. The overall LV complication-free rate was 98%. LV lead dislodgment occurred in 8 (1%) patients. PNS occurred in 58 (8%) patients, but only 3 (0.4%) patients required surgical intervention. CONCLUSION: The ACUITY X4 LV leads had low pacing thresholds particularly from proximal electrodes, a high incidence of pacing from the nondistal electrode, and low likelihood of dislodgment or PNS requiring surgical intervention. (ClinicalTrials.gov Identifier: NCT02071173).


Assuntos
Dispositivos de Terapia de Ressincronização Cardíaca , Terapia de Ressincronização Cardíaca , Desfibriladores Implantáveis , Cardioversão Elétrica/instrumentação , Insuficiência Cardíaca/terapia , Disfunção Ventricular Esquerda/terapia , Função Ventricular Esquerda , Idoso , Terapia de Ressincronização Cardíaca/efeitos adversos , Angiografia Coronária , Intervalo Livre de Doença , Cardioversão Elétrica/efeitos adversos , Feminino , Insuficiência Cardíaca/diagnóstico , Insuficiência Cardíaca/fisiopatologia , Humanos , Estimativa de Kaplan-Meier , Masculino , Pessoa de Meia-Idade , Estudos Prospectivos , Desenho de Prótese , Falha de Prótese , Fatores de Risco , Fatores de Tempo , Resultado do Tratamento , Estados Unidos , Disfunção Ventricular Esquerda/diagnóstico , Disfunção Ventricular Esquerda/fisiopatologia
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...