Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
J Chromatogr A ; 1208(1-2): 71-5, 2008 Oct 24.
Artigo em Inglês | MEDLINE | ID: mdl-18790482

RESUMO

To overcome the limitations of solid-phase extraction, we developed a device comprising a spin column packed with octadecyl silane-bonded monolithic silica for extracting amphetamines and methylenedioxyamphetamines from urine. Urine (0.5mL), buffer (0.4mL), and methoxyphenamine (internal standard) were directly put into the preactivated column. The column was centrifuged (3000rpm, 5min) for sample loading and washed. The adsorbed analytes were eluted and analyzed by high-performance liquid chromatography, without evaporation. The results were as follows: linear curves (drug concentrations of 0.2-20microg/mL); correlation coefficients >0.99; detection limit, 0.1microg/mL. The proposed method is not only useful for drugs from biological materials but also highly reproducible for the analysis of these drugs in urine.


Assuntos
Anfetaminas/urina , Métodos Analíticos de Preparação de Amostras , Cromatografia Líquida de Alta Pressão/métodos , Anfetaminas/isolamento & purificação , Concentração de Íons de Hidrogênio , Dióxido de Silício/química
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...