1.
J Pharm Sci ; 102(8): 2730-7, 2013 Aug.
Artigo em Inglês | MEDLINE | ID: mdl-23839955

RESUMO

Container closure integrity is a critical factor for maintaining product sterility and stability. Therefore, closure systems (found in vials, syringes, and cartridges) are designed to provide a seal between rubber stoppers and glass containers. To ensure that the contained product has maintained its sterility and stability at the time of deployment, the seal must remain intact within acceptable limits. To this end, a mathematical model has been developed to describe vacuum loss in evacuated drug vials. The model computes equivalent leak diameter corresponding to initial air leak rate as well as vacuum loss as a function of time and vial size. The theory accounts for three flow regimes that may be encountered. Initial leak rates from 10(-8) to 10(3) sccm (standard cubic centimeters per minute) were investigated for vials ranging from 1 to 100 mL. Corresponding leak diameters of 0.25-173 µm were predicted. The time for a vial to lose half of its vacuum, the T50 value, ranged from many years at the lowest leak rates and largest vials, to fractions of a second at the highest leak rates and smallest vials. These results may be used to determine what level of initial vacuum leak is acceptable for a given product.


Assuntos
Embalagem de Medicamentos , Ar/análise , Algoritmos , Embalagem de Medicamentos/métodos , Modelos Químicos , Vácuo