Rev Sci Instrum ; 92(4): 043529, 2021 Apr 01.
Artigo em Inglês | MEDLINE | ID: mdl-34243474

RESUMO

To efficiently determine the plasma electron density fluctuations using the MIR diagnostic technique, a 55-75 GHz 65 nm-CMOS transmitter has been developed where four separate intermediate frequency (IF) signals are up-converted, amplified, and then combined to generate an 8-tone RF output; a broadband 90 nm-CMOS receiver has also been constructed, which consists of an RF-low noise amplifier (LNA), mixer, and IF amplifier. The circuits and their corresponding modules will soon be deployed on the DIII-D and NSTX-U fusion devices. A 110-140 GHz 65 nm-CMOS receiver has also been designed, which is suitable for measuring the deep-core temperature fluctuations in the DIII-D tokamak using the electron cyclotron emission imaging diagnostic system. In addition to the RF-LNA/balun, mixer, and IF amplifier, an LO balun/tripler and driving amplifier are now included in this highly integrated circuit chip. By adopting the microwave and millimeter-wave system-on-chip concept in the front-end system design, this paper demonstrates that compact transmitter and receiver modules can be easily built, which, in turn, facilitates array implementation and maintenance.