Anal Chem ; 81(24): 10071-9, 2009 Dec 15.
Artigo em Inglês | MEDLINE | ID: mdl-19924863

RESUMO

Gas Chromatography (GC)-Mass Spectrometry (MS) with Atmospheric Pressure (AP) interface was introduced more than 30 years ago but never became a mainstream technique, mainly because of technical difficulties and cost of instrumentation. A recently introduced multipurpose AP source created the opportunity to reconsider the importance of AP ionization for GC. Here, we present an analytical evaluation of GC/APCI-MS showing the benefits of soft atmospheric pressure chemical ionization for GC in combination with a Time of Flight (TOF) mass analyzer. During this study, the complete analytical procedure was optimized and evaluated with respect to characteristic analytical parameters, such as repeatability, reproducibility, linearity, and detection limits. Limits of detection (LOD) were found within the range from 11.8 to 72.5 nM depending on the type of compound. The intraday and interday repeatability tests demonstrate relative standard deviations (RSDs) of peak areas between 0.7%-2.1% and 3.8%-6.4% correspondingly. Finally, we applied the developed method to the analysis of human cerebrospinal fluid (CSF) samples to check the potential of this new analytical combination for metabolic profiling.