1.
Guang Pu Xue Yu Guang Pu Fen Xi ; 34(7): 1954-8, 2014 Jul.
Artigo em Chinês | MEDLINE | ID: mdl-25269315

RESUMO

The purpose of this research is to study the influence of double pulse laser induced breakdown spectroscopy (DP-LIBS) on the sensitivity of Cu in water. The water solution of Cu was tested by collinear DP-LIBS in this article. The results show that spectral intensity of Cu can be enhanced obviously by DP-LIBS, compared with single pulse laser induced breakdown spectroscopy (SP-LIBS). Besides, the experimental results were significantly impacted by delay time between laser pulse and spectrometer acquisition, delay time of double laser pulse and energy of laser pulse and so on. The paper determined the best conditions for DP-LIBS detecting Cu in water. The optimal acquisition delay time was 1 380 ns. The best laser pulse delay time was 25 ns. The most appropriate energy of double laser pulse was 100 mJ. Characteristic analysis of spectra of Cu at 324.7 and 327.4 nm was done for quantitative analysis. The detection limit was 3.5 microg x mL(-1) at 324.7 nm, and the detection limit was 4.84 microg x mL(-1) at 327.4 nm. The relative standard deviation of the two characteristic spectral lines was within 10%. The calibration curve of characteristic spectral line, established by 327.4 nm, was verified with 500 microg x mL(-1) sample. Concentration of the sample was 446 microg x mL(-1) calculated by the calibration curve. This research shows that the detection sensitivity of Cu in water can be improved by DP-LIBS. At the same time, it had high stability.

2.
Guang Pu Xue Yu Guang Pu Fen Xi ; 33(11): 3120-3, 2013 Nov.
Artigo em Chinês | MEDLINE | ID: mdl-24555394

RESUMO

The present study is to improve the sensitivity of detection and reduce the limit of detection in detecting heavy metal of soil by laser induced breakdown spectroscopy (LIBS). The Cr element of national standard soil was regarded as the research object. In the experiment, a conical cavity with small diameter end of 20 mm and large diameter end of 45 mm respectively was installed below the focusing lens near the experiment sample to mainly confine the signal transmitted by plasma and to some extent to confine the plasma itself in the LIBS setup. In detecting Cr I 425.44 nm, the beast delay time gained from experiment is 1.3 micros, and the relative standard deviation is below 10%. Compared with the setup of non-spatial confinement, the spectral intensity of Cr in the soil sample was enhanced more than 7%. Calibration curve was established in the Cr concentration range from 60 to 400 microg x g(-1). Under the condition of spatial confinement, the liner regression coefficient and the limit of detection were 0.997 71 and 18.85 microg x g(-1) respectively, however, the regression coefficient and the limit of detection were 0.991 22 and 36.99 microg x g(-1) without spatial confinement. So, this shows that conical spatial confinement can/improve the sensitivity of detection and enhance the spectral intensity. And it is a good auxiliary function in detecting Cr in the soil by laser induced breakdown spectroscopy.

SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...