1.
PLoS One ; 15(9): e0239667, 2020.
Artigo em Inglês | MEDLINE | ID: mdl-32986751

RESUMO

Life expectancy of contemporary cardiac pacemakers is limited due to the use of an internal primary battery. Repeated device replacement interventions are necessary, which leads to an elevated risk for patients and an increase of health care costs. The aim of our study is to investigate the feasibility of powering an endocardial pacemaker by converting a minimal amount of the heart's kinetic energy into electric energy. The intrinsic cardiac muscle activity makes it an ideal candidate as continuous source of energy for endocardial pacemakers. For this reason, we developed a prototype able to generate enough power to supply a pacing circuit at different heart rates. The prototype consists of a mass imbalance that drives an electromagnetic generator while oscillating. We developed a mathematical model to estimate the amount of energy harvested from the right ventricle. Finally, the implemented prototype was successfully tested during in-vitro and in-vivo experiments.


Assuntos
Fontes de Energia Elétrica , Radiação Eletromagnética , Endocárdio/fisiologia , Desenho de Equipamento , Modelos Teóricos , Marca-Passo Artificial , Animais , Velocidade do Fluxo Sanguíneo , Estudos de Viabilidade , Frequência Cardíaca , Suínos , Função Ventricular Direita/fisiologia