1.
Appl Radiat Isot ; 202: 111066, 2023 Dec.
Artigo em Inglês | MEDLINE | ID: mdl-37865066

RESUMO

This study aims to report the relevant issues concerning small fields in the commissioning of a TrueBeam STx for photon energies of 6MV, 10MV, 6FFF, and 10FFF. Percent depth doses, profiles, and field output factors were measured according to the beam model configuration of the treatment planning system. Multiple detectors were used based on the IAEA TRS-483 protocol as well as EBT3 radiochromic film. Analytical Anisotropic and Acuros XB algorithms, were configured and validated through basic dosimetry comparisons and end-to-end clinical tests.


Assuntos
Radiometria , Planejamento da Radioterapia Assistida por Computador , Planejamento da Radioterapia Assistida por Computador/métodos , Radiometria/métodos , Algoritmos , Fótons/uso terapêutico , Carmustina
2.
Phys Eng Sci Med ; 44(4): 1273-1283, 2021 Dec.
Artigo em Inglês | MEDLINE | ID: mdl-34618329

RESUMO

Two methods for non-coplanar beam direction optimization, one for static beams and another for arc trajectories, were proposed for intracranial tumours. The results of the beam angle optimizations were compared with the beam directions used in the clinical plans. Ten meningioma cases already treated were selected for this retrospective planning study. Algorithms for non-coplanar beam angle optimization (BAO) and arc trajectory optimization (ATO) were used to generate the corresponding plans. A plan quality score, calculated by a graphical method for plan assessment and comparison, was used to guide the beam angle optimization process. For each patient, the clinical plans (CLIN), created with the static beam orientations used for treatment, and coplanar VMAT approximated plans (VMAT) were also generated. To make fair plan comparisons, all plan optimizations were performed in an automated multicriteria calculation engine and the dosimetric plan quality was assessed. BAO and ATO plans presented, on average, moderate global plan score improvements over VMAT and CLIN plans. Nevertheless, while BAO and CLIN plans assured a more efficient OARs sparing, the ATO and VMAT plans presented a higher coverage and conformity of the PTV. Globally, all plans presented high-quality dose distributions. No statistically significant quality differences were found, on average, between BAO, ATO and CLIN plans. However, automated plan solution optimizations (BAO or ATO) may improve plan generation efficiency and standardization. In some individual patients, plan quality improvements were achieved with ATO plans, demonstrating the possible benefits of this automated optimized delivery technique.


Assuntos
Neoplasias Meníngeas , Meningioma , Radioterapia de Intensidade Modulada , Humanos , Neoplasias Meníngeas/radioterapia , Meningioma/radioterapia , Dosagem Radioterapêutica , Planejamento da Radioterapia Assistida por Computador , Estudos Retrospectivos
3.
Radiat Oncol ; 10: 41, 2015 Feb 15.
Artigo em Inglês | MEDLINE | ID: mdl-25889988

RESUMO

PURPOSE: To evaluate the differences between three methods of classification of recurrences in patients with head and neck tumours treated with Radiation Therapy (RT). MATERIALS AND METHODS: 367 patients with head and neck tumours were included in the study. Tumour recurrences were delineated in the CT images taken during patient follow-up and deformable registration was used to transfer this volume into the planning CT. The methods used to classify recurrences were: method CTV quantified the intersection volume between the recurrence and the Clinical Target Volume (CTV); method TV quantified the intersection between the Treated Volume and the recurrence (for method CTV and TV, recurrences were classified in-field if more than 95% of their volume were inside the volume of interest, marginal if the intersection was between 20-95% and outfield otherwise); and method COM was based on the position of the Centre Of Mass of the recurrence. A dose assessment in the recurrence volume was also made. RESULTS: The 2-year Kaplan-Meier locoregional recurrence incidence was 10%. Tumour recurrences occurred in 22 patients in a mean time of 16.5 ± 9.4 months resulting in 28 recurrence volumes. The percentage of in-field recurrences for methods CTV, TV and COM was 7%, 43% and 50%, respectively. Agreement between the three methods in characterizing individually in-field and marginal recurrences was found only in six cases. Methods CTV and COM agreed in 14. The percentage of outfield recurrences was 29% using all methods. For local recurrences (in-field or marginal to gross disease) the average difference between the prescribed dose and D 98% in the recurrence volume was -5.2 ± 3.5% (range: -10.1%-0.9%). CONCLUSIONS: The classification of in-field and marginal recurrences is very dependent on the method used to characterize recurrences. Using methods TV and COM the largest percentage of tumour recurrences occurred in-field in tissues irradiated with high doses.


Assuntos
Carcinoma de Células Escamosas/radioterapia , Neoplasias de Cabeça e Pescoço/radioterapia , Recidiva Local de Neoplasia/classificação , Recidiva Local de Neoplasia/diagnóstico , Radioterapia de Intensidade Modulada/métodos , Radioterapia/efeitos adversos , Adolescente , Adulto , Idoso , Idoso de 80 Anos ou mais , Carcinoma de Células Escamosas/mortalidade , Carcinoma de Células Escamosas/patologia , Criança , Feminino , Seguimentos , Neoplasias de Cabeça e Pescoço/mortalidade , Neoplasias de Cabeça e Pescoço/patologia , Humanos , Masculino , Pessoa de Meia-Idade , Recidiva Local de Neoplasia/etiologia , Estadiamento de Neoplasias , Prognóstico , Planejamento da Radioterapia Assistida por Computador/métodos , Estudos Retrospectivos , Medição de Risco , Taxa de Sobrevida , Tomografia Computadorizada por Raios X/métodos , Adulto Jovem
4.
Acta Oncol ; 53(5): 628-36, 2014 May.
Artigo em Inglês | MEDLINE | ID: mdl-24164104

RESUMO

BACKGROUND AND PURPOSE: One of the newer audit modalities operated by the International Atomic Energy Agency (IAEA) involves audits of treatment planning systems (TPS) in radiotherapy. The main focus of the audit is the dosimetry verification of the delivery of a radiation treatment plan for three-dimensional (3D) conformal radiotherapy using high energy photon beams. The audit has been carried out in eight European countries - Estonia, Hungary, Latvia, Lithuania, Serbia, Slovakia, Poland and Portugal. The corresponding results are presented. MATERIAL AND METHODS: The TPS audit reviews the dosimetry, treatment planning and radiotherapy delivery processes using the 'end-to-end' approach, i.e. following the pathway similar to that of the patient, through imaging, treatment planning and dose delivery. The audit is implemented at the national level with IAEA assistance. The national counterparts conduct the TPS audit at local radiotherapy centres through on-site visits. TPS calculated doses are compared with ion chamber measurements performed in an anthropomorphic phantom for eight test cases per algorithm/beam. A set of pre-defined agreement criteria is used to analyse the performance of TPSs. RESULTS: TPS audit was carried out in 60 radiotherapy centres. In total, 190 data sets (combination of algorithm and beam quality) have been collected and reviewed. Dosimetry problems requiring interventions were discovered in about 10% of datasets. In addition, suboptimal beam modelling in TPSs was discovered in a number of cases. CONCLUSIONS: The TPS audit project using the IAEA methodology has verified the treatment planning system calculations for 3D conformal radiotherapy in a group of radiotherapy centres in Europe. It contributed to achieving better understanding of the performance of TPSs and helped to resolve issues related to imaging, dosimetry and treatment planning.


Assuntos
Garantia da Qualidade dos Cuidados de Saúde , Radiometria/normas , Planejamento da Radioterapia Assistida por Computador/normas , Radioterapia Conformacional/normas , Europa (Continente) , Humanos , Agências Internacionais , Auditoria Médica
5.
Radiat Oncol ; 5: 57, 2010 Jun 22.
Artigo em Inglês | MEDLINE | ID: mdl-20569482

RESUMO

PURPOSE: To quantify the radiobiological advantages obtained by an Improved Forward Planning technique (IFP) and two IMRT techniques using different fractionation schemes for the irradiation of head and neck tumours. The conventional radiation therapy technique (CONVT) was used here as a benchmark. METHODS: Seven patients with head and neck tumours were selected for this retrospective planning study. The PTV1 included the primary tumour, PTV2 the high risk lymph nodes and PTV3 the low risk lymph nodes. Except for the conventional technique where a maximum dose of 64.8 Gy was prescribed to the PTV1, 70.2 Gy, 59.4 Gy and 50.4 Gy were prescribed respectively to PTV1, PTV2 and PTV3. Except for IMRT2, all techniques were delivered by three sequential phases. The IFP technique used five to seven directions with a total of 15 to 21 beams. The IMRT techniques used five to nine directions and around 80 segments. The first, IMRT1, was prescribed with the conventional fractionation scheme of 1.8 Gy per fraction delivered in 39 fractions by three treatment phases. The second, IMRT2, simultaneously irradiated the PTV2 and PTV3 with 59.4 Gy and 50.4 Gy in 28 fractions, respectively, while the PTV1 was boosted with six subsequent fractions of 1.8 Gy. Tissue response was calculated using the relative seriality model and the Poisson Linear-Quadratic-Time model to simulate repopulation in the primary tumour. RESULTS: The average probability of total tumour control increased from 38% with CONVT to 80% with IFP, to 85% with IMRT1 and 89% with IMRT2. The shorter treatment time and larger dose per fraction obtained with IMRT2 resulted in an 11% increase in the probability of control in the PTV1 with respect to IFP and 7% relatively to IMRT1 (p < 0.05). The average probability of total patient complications was reduced from 80% with CONVT to 61% with IFP and 31% with IMRT. The corresponding probability of complications in the ipsilateral parotid was 63%, 42% and 20%; in the contralateral parotid it was 50%, 20% and 9%; in the oral cavity it was 2%, 15% and 4% and in the mandible it was 1%, 5% and 3%, respectively. CONCLUSIONS: A significant improvement in treatment outcome was obtained with IMRT compared to conventional radiation therapy. The practical and biological advantages of IMRT2, employing a shorter treatment time, may outweigh the small differences obtained in the organs at risk between the two IMRT techniques. This technique is therefore presently being used in the clinic for selected patients with head and neck tumours. A significant improvement in the quality of the dose distribution was obtained with IFP compared to CONVT. Thus, this beam arrangement is used in the clinical routine as an alternative to IMRT.