Clin Chim Acta ; 89(1): 79-91, 1978 Oct 02.
Artigo em Inglês | MEDLINE | ID: mdl-30551

RESUMO

A new automated method is proposed for the accurate determination of the true creatinine concentration in 20 microliter serum, based on the measurement of the initial rate of the decomposition reaction of the creatinine picrate complex caused by decreasing the pH. The results of this method, performed with a Centrifichem centrifugal analyzer system (type 300 F), are compared with the results obtained by two other methods: a manual method based upon the adsorption on Fuller's earth, and the Auto Analyzer method. In these comparative studies special attention is paid to the analysis of sera containing low (normal) substrate concentrations. Calculation of the orthogonal regression between the results obtained by the adsorption method (x) and the proposed method (y) gave: y=0.97x +/- 5 mumol/l (x=91.3 mumol/l, y-93.5 mumol/l), while for the correlation coefficient (r) 0.9721 was found. All sera (n=113) had creatinine concentrations between 40 and 180 mumol/l.


Assuntos
Creatinina/sangue , Centrifugação/instrumentação , Creatinina/urina , Liofilização , Humanos , Concentração de Íons de Hidrogênio , Métodos , Fatores de Tempo