Artigo em Coreano | WPRIM (Pacífico Ocidental) | ID: wpr-196927

RESUMO

PURPOSE: To compare electroretinogram (ERG) waveforms acquired using an ERG-jet electrode and a Burian-Allen electrode. METHODS: ERGs were recorded with an ERG-jet electrode and a Burian-Allen electrode from both eyes of 29 volunteers. Three consecutive recordings were obtained with both electrodes from the left eye of another 6 volunteers. Peak-to-trough amplitudes and peak implicit times were compared between both eyes, and between the 2 types of electrodes. Interpersonal and intrapersonal variation were also compared. RESULTS: The differences in amplitude and implicit time between the right and left eyes were not influenced by the type of electrode. The amplitude of the ERG acquired using the Burian-Allen electrode, however, showed significant differences between both eyes, contrary to the amplitude acquired using the ERG-jet electrode. ERG recordings obtained using the ERG-jet electrode generally showed higher amplitude, shorter implicit time, and less interpersonal and intrapersonal variation than ERG recordings obtained using the Burian-Allen electrode. CONCLUSIONS: In this application, the ERG-jet ERG electrode appeared to be superior to the Burian-Allen electrode in terms of both usefulness and consistency.


Assuntos
Eletrodos , Olho
2.
Artigo em Coreano | WPRIM (Pacífico Ocidental) | ID: wpr-120421

RESUMO

Using Burian-Allen bipolar contact lens electrode, Authors examined the normal 120 eyes of 120 subjects(male 57, female 63), age ranged 7 to 83 years(mean 44.5 years), with the method which was same as the ISCEV standard. There was no statistically significant differences between the value obtained using Burian-Allen electrode and the value obtained using ERG-jet electrode(p>0.05). We expect that this data would be useful testing method to evaluate the retinal disease. And this data may provide new reference value in the study of electroretinogram.


Assuntos
Feminino , Humanos , Eletrodos , Eletrorretinografia , Valores de Referência , Doenças Retinianas
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...