1.
J Cardiovasc Electrophysiol ; 35(4): 794-801, 2024 Apr.
Artigo em Inglês | MEDLINE | ID: mdl-38384108

RESUMO

INTRODUCTION: Several implantable cardioverter defibrillators (ICD) programming strategies are applied to minimize ICD therapy, especially unnecessary therapies from supraventricular arrhythmias (SVA). However, it remains unknown whether these optimal programming recommendations only benefit those with SVAs or have any detrimental effects from delayed therapy on those without SVAs. This study aims to assess the impact of SVA on the outcomes of ICD programming based on 2015 HRS/EHRA/APHRS/SOLAECE expert consensus statement and 2019 focused update on optimal ICD programming and testing guidelines. METHODS: Consecutive patients who underwent ICD insertion for primary prevention were classified into four groups based on SVA status and ICD programming: (1) guideline-concordant group (GC) with SVA, (2) GC without SVA, (3) nonguideline concordant group (NGC) with SVA, and (4) NGC without SVA. Cox proportional hazard models were analyzed for freedom from ICD therapies, shock, and mortality. RESULTS: Seven hundred and seventy-two patients (median age, 64 years) were enrolled. ICD therapies were the most frequent in NGC with SVA (24.0%), followed by NGC without SVA (19.9%), GC without SVA (11.6%), and GC with SVA (8.1%). Guideline concordant programming was associated with 68% ICD therapy reduction (HR 0.32, p = .007) and 67% ICD shock reduction (HR 0.33, p = .030) in SVA patients and 44% ICD therapy reduction in those without SVA (HR 0.56, p = .030). CONCLUSION: Programming ICDs in primary prevention patients based on current guidelines reduces therapy burden without increasing mortality in both SVA and non-SVA patients. A greater magnitude of reduced ICD therapy was found in those with supraventricular arrhythmias.


Assuntos
Desfibriladores Implantáveis , Humanos , Pessoa de Meia-Idade , Desfibriladores Implantáveis/efeitos adversos , Cardioversão Elétrica/efeitos adversos , Arritmias Cardíacas , Morte Súbita Cardíaca/prevenção & controle
2.
Artif Organs ; 48(5): 525-535, 2024 May.
Artigo em Inglês | MEDLINE | ID: mdl-38213270

RESUMO

BACKGROUND: Left ventricular assist device (L-VAD) implantation is increasingly used in patients with heart failure (HF) and most patients also have an implantable cardioverter defibrillator (ICD). Limited data are available on the incidence of ICD therapies and complications in this special setting. The aim of this study was to analyze the real-world incidence and predictors of ICD therapies, complications and interactions between ICD and L-VAD. METHODS: We conducted a multicenter retrospective observational study in patients with advanced HF implanted with ICD and a continuous-flow L-VAD, followed-up in five advanced HF centers in Northern Italy. RESULTS: A total of 234 patients (89.7% male, median age 59, 48.3% with ischemic etiology) were enrolled. After a median follow-up of 21 months, 66 patients (28.2%) experienced an appropriate ICD therapy, 22 patients (9.4%) an inappropriate ICD therapy, and 17 patients (7.3%) suffered from an interaction between ICD and L-VAD. The composite outcome of all ICD-related complications was reported in 41 patients (17.5%), and 121 (51.7%) experienced an L-VAD-related complication. At multivariable analysis, an active ventricular tachycardia (VT) zone and a prior ICD generator replacement were independent predictors of ICD therapies and of total ICD-related complications, respectively. CONCLUSIONS: Real-world patients with both L-VAD and ICD experience a high rate of ICD therapies and complications. Our findings suggest the importance of tailoring device programming in order to minimize the incidence of unnecessary ICD therapies, thus sparing the need for ICD generator replacement, a procedure associated to a high risk of complications.


Assuntos
Desfibriladores Implantáveis , Insuficiência Cardíaca , Coração Auxiliar , Taquicardia Ventricular , Feminino , Humanos , Masculino , Arritmias Cardíacas/etiologia , Desfibriladores Implantáveis/efeitos adversos , Insuficiência Cardíaca/cirurgia , Insuficiência Cardíaca/etiologia , Coração Auxiliar/efeitos adversos , Estudos Retrospectivos , Taquicardia Ventricular/etiologia , Resultado do Tratamento , Pessoa de Meia-Idade
3.
Am J Cardiovasc Dis ; 13(4): 207-221, 2023.
Artigo em Inglês | MEDLINE | ID: mdl-37736352

RESUMO

Enhanced ventricular arrhythmogenesis is commonly experienced by patients in the end-stage of heart failure spectrum. A high burden of ventricular arrhythmias can affect the ventricular systolic function, lead to unexpected hospitalizations and further deteriorate the prognosis. Management of ventricular arrhythmias in this population is challenging. Implantable cardioverter-defibrillators are protective for the immediate termination of life-threatening arrhythmias but they have no impact in reducing the arrhythmic burden. Combination treatment with invasive (catheter ablation, mechanical hemodynamic support, sympathetic denervation) and noninvasive (antiarrhythmic drugs, medical therapy for heart failure, programming of implantable devices) therapies is commonly required. The aim of this review is to present the available therapeutic options, with main focus on recently published data for catheter ablation and provide a stepwise treatment approach.

4.
J Interv Card Electrophysiol ; 66(6): 1359-1366, 2023 Sep.
Artigo em Inglês | MEDLINE | ID: mdl-36422768

RESUMO

BACKGROUND: While ICD therapy reduction programming strategies are recommended in current multi-society guidelines, concerns remain about a possible trade-off between the benefits of ICD therapy reduction and failure to treat episodes of ventricular arrhythmias. The study is to evaluate the outcomes of primary prevention patients followed in centers with high and low concordance with the 2015 HRS/EHRA/APHRS/SOLAECE expert consensus statement and 2019 focused update on optimal ICD programming and testing guidelines. METHODS: Consecutive patients with primary prevention ICD implantation from two centers between 2014 and 2016 were included. One center was classified as high guideline concordance center (HGC) with 47% (146/310) of patients with initial ICD concordant with the guidelines, and the other center was classified as low guideline concordance center (LGC) with only 1% (2/178) of patients with guideline-concordant initial ICD programming. Cox proportional hazard models were used to assess risk of first ICD therapy (ATP or shock), first ICD shock, and mortality. RESULTS: A total of 488 patients were included (mean age, 66 ± 13 years). During a mean follow-up of 1.9 ± 0.9 years, patients followed at HGC were 63% less likely to receive any ICD therapy (adjusted HR [aHR] 0.37, 95% CI 0.42-0.99). There were no significant differences in the rate of first ICD shock (aHR 0.72, 95% CI 0.34-1.52) or mortality (aHR 1.19, 95% CI, 0.47-3.05). CONCLUSIONS: Compared to primary prevention patients followed at LGC, primary prevention ICD patients followed at HGC received a significantly lower rate of ICD therapy, mainly from ATP reduction, without a difference in mortality during follow-up.


Assuntos
Desfibriladores Implantáveis , Humanos , Pessoa de Meia-Idade , Idoso , Desfibriladores Implantáveis/efeitos adversos , Cardioversão Elétrica/efeitos adversos , Arritmias Cardíacas/terapia , Trifosfato de Adenosina , Prevenção Primária , Morte Súbita Cardíaca/prevenção & controle , Morte Súbita Cardíaca/etiologia
6.
Front Cardiovasc Med ; 9: 1011619, 2022.
Artigo em Inglês | MEDLINE | ID: mdl-36684577

RESUMO

Management of hemodynamically stable, incessant wide QRS complex tachycardia (WCT) in patients who already have an implantable cardioverter defibrillator (ICD) is challenging. First-line treatment is performed by medical staff who have no knowledge on programmed ICD therapy settings and there is always some concern for unexpected ICD shock. In these patients, a structured approach is necessary from presentation to therapy. The present review provides a systematic approach in four distinct phases to guide any physician involved in the management of these patients: PHASE I: assessment of hemodynamic status and use of the magnet to temporarily suspend ICD therapies, especially shocks; identification of possible arrhythmia triggers; risk stratification in case of electrical storm (ES). PHASE II: The preparation phase includes reversal of potential arrhythmia "triggers", mild patient sedation, and patient monitoring for therapy delivery. Based on resource availability and competences, the most adequate therapeutic approach is chosen. This choice depends on whether a device specialist is readily available or not. In the case of ES in a "high-risk" patient an accelerated patient management protocol is advocated, which considers urgent ventricular tachycardia transcatheter ablation with or without mechanical cardiocirculatory support. PHASE III: Therapeutic phase is based on the use of intravenous anti-arrhythmic drugs mostly indicated in this clinical context are presented. Device interrogation is very important in this phase when sustained monomorphic VT diagnosis is confirmed, then ICD ATP algorithms, based on underlying VT cycle length, are proposed. In high-risk patients with intractable ES, intensive patient management considers MCS and transcatheter ablation. PHASE IV: The patient is hospitalized for further diagnostics and management aimed at preventing arrhythmia recurrences.

7.
Heart Rhythm ; 18(8): 1344-1351, 2021 08.
Artigo em Inglês | MEDLINE | ID: mdl-33887451

RESUMO

BACKGROUND: The role of cardiovascular implantable electronic device (CIED)-derived activity to predict implantable cardioverter-defibrillator (ICD) therapy or death is not known. OBJECTIVE: We aimed to assess CIED-derived activity to predict clinical outcomes. METHODS: In 1500 patients enrolled in MADIT-RIT, CIED-derived patient activity was acquired daily, then averaged for the first 30 days following randomization to predict inappropriate/appropriate therapy or death. Kaplan-Meier analysis and Cox proportional regression models were used to evaluate inappropriate/appropriate therapy, heart failure, or death by 30-day CIED-derived patient activity quintiles. RESULTS: There were 1463 patients with CIED activity data (98%). Patients in the highest quintile (Q5) of activity (more active) had the highest rate of inappropriate therapy, 21% at 2 years, as compared to 7%-11% in the other 4 quintiles (P < .001), a 1.75 times higher risk (95% confidence interval [CI]: 1.23-2.50, P = .002). However, patients in the lowest quintile of activity (Q1, 1 hour/day) had the highest risk of mortality, 15% in 2 years, as compared to Q2-3 (1-2 hours/day, 8%-7% mortality), and Q4-5 (>2 hours/day, 2%-3% mortality) (P < .001). Patients with the lowest level of activity (Q1) had a 2.02 times higher risk of mortality (95% CI: 1.21-3.38, P = .007), and they had an 82% higher risk of heart failure hospitalization (95% CI: 1.28-2.57, P = .001). CONCLUSIONS: High CIED-derived 30-day median patient activity predicted inappropriate therapy, while low patient activity predicted mortality and heart failure in ICD and cardiac resynchronization therapy with defibrillator patients enrolled in MADIT-RIT. Device-derived activity assessment could serve as a useful predictor of outcomes.


Assuntos
Desfibriladores Implantáveis/efeitos adversos , Insuficiência Cardíaca/etiologia , Taquicardia Ventricular/terapia , Idoso , Falha de Equipamento , Feminino , Saúde Global , Insuficiência Cardíaca/epidemiologia , Humanos , Incidência , Masculino , Pessoa de Meia-Idade , Prognóstico , Fatores de Risco , Taxa de Sobrevida/tendências
8.
Rev. esp. cardiol. (Ed. impr.) ; 74(4): 296-302, Abr. 2021. tab, graf
Artigo em Espanhol | IBECS | ID: ibc-232234

RESUMO

Introducción y objetivos El estudio ADVANCE III mostró una reducción de terapias del desfibrilador automático implantable (DAI) con el empleo de tiempos de detección de arritmia prolongados. Se describe la adopción y el impacto de dicha estrategia en la incidencia de terapias del DAI. Métodos Estudio observacional retrospectivo de pacientes con implante de DAI Medtronic (2005-2016) en un registro multicéntrico (UMBRELLA-NCT01561144). Se describe la evolución de la adopción de programación ADVANCE en relación con: publicación del estudio, implementación de una campaña de formación y publicación de un consenso de expertos. Se identificaron con regresión logística los predictores de la adopción. Se comparó la incidencia de terapias en pacientes con y sin programación ADVANCE estimando la razón de tasas de incidencia ajustada (RTIa) mediante regresión binomial negativa. Resultados Se incluyó a 3.528 pacientes. Se utilizó la estrategia ADVANCE en el 20% del total y el 44% al final del estudio. La adopción se incrementó tras publicarse el estudio ADVANCE, y en menor grado tras la campaña de formación y consenso de expertos. Predictores de la adopción: DAI con detección nominal 30/40 (ORa=4,4; IC95%, 3,5-5,4), implantador electrofisiólogo (ORa=1,7; IC95%, 1,4-2,2) y prevención secundaria (ORa=3,2; IC95%, 2,6-3,9). El implante de DAI bicameral (ORa=0,6; IC95%, 0,5-0,8) o tricameral (ORa=0,5; IC95%, 0,4-0,7) se asoció con menor adopción. La programación ADVANCE se asoció con reducción de terapias totales (RTIa=0,77; IC95%, 0,69-0,86) y choques inapropiados (RTIa=0,66; IC95%, 0,52-0,85). Conclusiones La adopción de la programación ADVANCE es poco amplia y puede mejorarse mediante una adecuada selección de los parámetros nominales. Emplearla se asocia con una reducción de las terapias del DAI. (AU)


Introduction and objectives The ADVANCE III trial showed that a delayed-detection strategy reduces implantable cardioverter-defibrillator (ICD) therapies. Here, we describe the adherence to and predictors of ADVANCE adoption and compare ICD therapy rates between patients with and without ADVANCE programming. Methods This observational retrospective study analyzed patients implanted with Medtronic ICDs included from 2005 to 2016 in a Spanish national multicenter registry (UMBRELLA database; ClinicalTrials.gov, NCT01561144). Changes in ADVANCE programming adoption were described in relation to a) publication of the ADVANCE trial, b) implementation of an “ADVANCE awareness” campaign, and c) publication of an expert consensus statement. Multivariate logistic regression identified predictors of adoption. Therapy incidence rates were compared between groups by estimating the adjusted incidence rate ratio (aIRR) using negative binomial regression. Results A total of 3528 patients were included. An ADVANCE strategy was used in 20% overall and in 44% at the end of the study. ADVANCE III adoption increased after trial publication, with less growth after an “ADVANCE awareness” campaign and after expert consensus statement publication. Predictors of ADVANCE adoption were as follows: ICD device with a nominal number of intervals to detect 30/40 (aOR, 4.4; 95%CI, 3.5-5.4), implantation by an electrophysiologist (aOR, 1.7; 95%CI, 1.4-2.2), and secondary prevention (aOR, 3.2; 95%CI, 2.6-3.9). Dual-chamber ICDs (aOR, 0.6; 95%CI, 0.5-0.8) and cardiac resynchronization-defibrillators (aOR, 0.5; 95%CI, 0.4-0.7) were associated with lower adoption. ADVANCE programming was associated with reduced total therapy burden (aIRR, 0.77; 95%CI, 0.69-0.86) and fewer inappropriate shocks (aIRR, 0.66; 95%CI, 0.52-0.85). Conclusions ADVANCE adoption remains modest and can be improved through evidence-driven selection of nominal ICD settings. ... (AU)


Assuntos
Humanos , Masculino , Feminino , Pessoa de Meia-Idade , Idoso , Desfibriladores Implantáveis , Medicina de Precisão , Estudos Retrospectivos , Epidemiologia Descritiva
9.
Rev Esp Cardiol (Engl Ed) ; 74(4): 296-302, 2021 Apr.
Artigo em Inglês, Espanhol | MEDLINE | ID: mdl-32773348

RESUMO

INTRODUCTION AND OBJECTIVES: The ADVANCE III trial showed that a delayed-detection strategy reduces implantable cardioverter-defibrillator (ICD) therapies. Here, we describe the adherence to and predictors of ADVANCE adoption and compare ICD therapy rates between patients with and without ADVANCE programming. METHODS: This observational retrospective study analyzed patients implanted with Medtronic ICDs included from 2005 to 2016 in a Spanish national multicenter registry (UMBRELLA database; ClinicalTrials.gov, NCT01561144). Changes in ADVANCE programming adoption were described in relation to a) publication of the ADVANCE trial, b) implementation of an "ADVANCE awareness" campaign, and c) publication of an expert consensus statement. Multivariate logistic regression identified predictors of adoption. Therapy incidence rates were compared between groups by estimating the adjusted incidence rate ratio (aIRR) using negative binomial regression. RESULTS: A total of 3528 patients were included. An ADVANCE strategy was used in 20% overall and in 44% at the end of the study. ADVANCE III adoption increased after trial publication, with less growth after an "ADVANCE awareness" campaign and after expert consensus statement publication. Predictors of ADVANCE adoption were as follows: ICD device with a nominal number of intervals to detect 30/40 (aOR, 4.4; 95%CI, 3.5-5.4), implantation by an electrophysiologist (aOR, 1.7; 95%CI, 1.4-2.2), and secondary prevention (aOR, 3.2; 95%CI, 2.6-3.9). Dual-chamber ICDs (aOR, 0.6; 95%CI, 0.5-0.8) and cardiac resynchronization-defibrillators (aOR, 0.5; 95%CI, 0.4-0.7) were associated with lower adoption. ADVANCE programming was associated with reduced total therapy burden (aIRR, 0.77; 95%CI, 0.69-0.86) and fewer inappropriate shocks (aIRR, 0.66; 95%CI, 0.52-0.85). CONCLUSIONS: ADVANCE adoption remains modest and can be improved through evidence-driven selection of nominal ICD settings. ADVANCE programming is associated with reduced therapy rates in real-world ICD recipients.


Assuntos
Terapia de Ressincronização Cardíaca , Desfibriladores Implantáveis , Cardioversão Elétrica , Humanos , Incidência , Estudos Retrospectivos
11.
J Cardiovasc Electrophysiol ; 29(10): 1418-1424, 2018 10.
Artigo em Inglês | MEDLINE | ID: mdl-29978932

RESUMO

INTRODUCTION: Data on inappropriate and appropriate ICD therapy, and efficacy of ICD programing strategies by race are limited. METHODS: In MADIT-RIT, we evaluated the risk of ICD therapy by race, and the efficacy of high rate cut-off ventricular tachycardia (VT) zone ≥200 beats per minute (bpm) (Arm B), or 60 seconds delay in VT zone 170-199 bpm (Arm C), compared to 2.5 seconds delay at 170 bpm (Arm A) among black and white patients. RESULTS: MADIT-RIT enrolled 272 (20%) black and 1119 (80%) white patients. The risk of inappropriate therapy was similar among blacks and whites, HR 1.25, 95% CI (0.82-1.93), P  =  0.30. High rate cut-off or delayed VT therapy was associated with significant reductions in inappropriate therapy among whites, Arm B versus Arm A, HR 0.15, 95% CI (0.08-0.29), P < 0.0001, Arm C versus Arm A, HR 0.19, 95% CI (0.11-0.33), P < 0.001, and black individuals Arm B versus Arm A, HR 0.24, 95% CI (0.01-0.56), P  =  0.0001, Arm C versus Arm A, HR 0.30, 95% CI (0.13-0.68), P  =  0.004, P interaction > 0.10). However, delayed VT therapy was associated with a trend toward greater reduction in appropriate therapy in black individuals, HR 0.08, 95% CI (0.03-0.27), P < 0.0001 relative to white individuals, HR 0.27, 95% CI (0.16-0.43), P < 0.0001, P interaction  =  0.077. CONCLUSION: In MADIT-RIT, high rate and delayed detection ICD programming provided similar benefit with reductions in both inappropriate therapy and unnecessary appropriate therapy among black and white individuals. CLINICALTRIALS. GOV IDENTIFIER: NCT00947310.


Assuntos
População Negra , Desfibriladores Implantáveis , Cardioversão Elétrica/instrumentação , Disparidades em Assistência à Saúde/etnologia , Taquicardia Ventricular/terapia , População Branca , Potenciais de Ação , Negro ou Afro-Americano , Cardioversão Elétrica/efeitos adversos , Europa (Continente)/epidemiologia , Feminino , Frequência Cardíaca , Humanos , Israel/epidemiologia , Japão/epidemiologia , Masculino , Pessoa de Meia-Idade , Estudos Multicêntricos como Assunto , América do Norte/epidemiologia , Desenho de Prótese , Falha de Prótese , Ensaios Clínicos Controlados Aleatórios como Assunto , Estudos Retrospectivos , Medição de Risco , Fatores de Risco , Taquicardia Ventricular/diagnóstico , Taquicardia Ventricular/etnologia , Taquicardia Ventricular/fisiopatologia , Fatores de Tempo , Resultado do Tratamento , Procedimentos Desnecessários
12.
Pacing Clin Electrophysiol ; 40(12): 1405-1411, 2017 Dec.
Artigo em Inglês | MEDLINE | ID: mdl-29052236

RESUMO

BACKGROUND: The effects of heart failure (HF) severity on risk of inappropriate implantable cardioverter-defibrillator (ICD) therapy have not been thoroughly investigated. We aimed to study the association between HF severity and inappropriate ICD therapy in MADIT-RIT. METHODS: MADIT-RIT randomized 1,500 patients to three ICD programming arms: conventional (Arm A), high-rate cut-off (Arm B: ≥200 beats/min), and delayed therapy (Arm C: 60-second delay for ≥170 beats/min). We evaluated the association between New York Heart Association (NYHA) class III (n = 256) versus class I-II (n = 251) and inappropriate ICD therapy in Arm A patients with ICD-only and cardiac resynchronization therapy with defibrillator (CRT-D). We additionally assessed benefit of novel ICD programming in Arms B and C versus Arm A by NYHA classification. RESULTS: In Arm A, the risk of inappropriate therapy was significantly higher in those with NYHA III versus NYHA I-II for both ICD (hazard ratio [HR] = 2.55, confidence interval [CI]: 1.51-4.30, P < 0.001) and CRT-D patients (HR = 3.73, CI: 1.14-12.23, P = 0.030). This was consistent for inappropriate ATP and inappropriate ICD therapy < 200 beats/min, but not for inappropriate shocks. Novel ICD programming significantly reduced inappropriate therapy in patients with both NYHA III (Arm B vs Arm A: HR = 0.08, P < 0.001; Arm C vs Arm A: HR = 0.17, P < 0.001) and NYHA I-II (Arm B vs Arm A: HR = 0.25, P < 0.001; Arm C vs Arm A: HR = 0.28, P < 0.001). CONCLUSION: Patients with more severe HF are at increased risk for inappropriate ICD therapy, particularly ATP due to arrhythmias < 200 beats/min. Novel programming with high-rate cut-off or delayed detection reduces inappropriate ICD therapies in both mild and moderate HF.


Assuntos
Terapia de Ressincronização Cardíaca , Desfibriladores Implantáveis , Insuficiência Cardíaca/terapia , Feminino , Humanos , Masculino , Uso Excessivo dos Serviços de Saúde , Pessoa de Meia-Idade , Ensaios Clínicos Controlados Aleatórios como Assunto , Estudos Retrospectivos , Índice de Gravidade de Doença
13.
Circ Arrhythm Electrophysiol ; 9(1): e001965, 2016 Jan.
Artigo em Inglês | MEDLINE | ID: mdl-26743237

RESUMO

BACKGROUND: The Multicenter Automatic Defibrillator Implantation Trial-Reduce Inappropriate therapy (MADIT-RIT) trial showed a significant reduction in inappropriate implantable cardioverter defibrillator (ICD) therapy in patients programmed to high-rate cut-off (Arm B) or delayed ventricular tachycardia therapy (Arm C), compared with conventional programming (Arm A). There is limited data on the effect of cardiac resynchronization therapy with a cardioverter defibrillator (CRT-D) on the effect of ICD programming. We aimed to elucidate the effect of CRT-D on ICD programming to reduce inappropriate ICD therapy in patients implanted with CRT-D or an ICD, enrolled in MADIT-RIT. METHODS AND RESULTS: The primary end point of this study was the first inappropriate ICD therapy. Secondary end points were inappropriate anti-tachycardia pacing and inappropriate ICD shock. The study enrolled 742 (49%) patients with an ICD and 757 (51%) patients with a CRT-D. Patients implanted with a CRT-D had 62% lower risk of inappropriate ICD therapy than those with an ICD only (hazard ratio [HR] =0.38, 95% confidence interval: 0.25-0.57; P<0.001). High-rate cut-off or delayed ventricular tachycardia therapy programming significantly reduced the risk of inappropriate ICD therapy compared with conventional ICD programming in ICD (HR=0.14 [B versus A]; HR=0.21 [C versus A]) and CRT-D patients (HR=0.15 [B versus A]; HR=0.23 [C versus A]; P<0.001 for all). There was a significant reduction in inappropriate anti-tachycardia pacings in both group and a significant reduction in inappropriate ICD shock in CRT-D patients. CONCLUSIONS: Patients implanted with a CRT-D have lower risk of inappropriate ICD therapy than those with an ICD. Innovative ICD programming significantly reduces the risk of inappropriate ICD therapy in both ICD and CRT-D patients. CLINICAL TRIAL REGISTRATION: http://clinicaltrials.gov; Unique identifier: NCT00947310.


Assuntos
Desfibriladores Implantáveis , Cardioversão Elétrica/instrumentação , Taquicardia Supraventricular/terapia , Taquicardia Ventricular/terapia , Idoso , Canadá/epidemiologia , Cardioversão Elétrica/efeitos adversos , Falha de Equipamento , Europa (Continente)/epidemiologia , Feminino , Seguimentos , Humanos , Japão/epidemiologia , Estimativa de Kaplan-Meier , Masculino , Pessoa de Meia-Idade , Fatores de Risco , Taxa de Sobrevida/tendências , Taquicardia Supraventricular/fisiopatologia , Taquicardia Ventricular/mortalidade , Taquicardia Ventricular/fisiopatologia , Fatores de Tempo , Resultado do Tratamento , Estados Unidos/epidemiologia
14.
Expert Rev Cardiovasc Ther ; 14(3): 291-9, 2016.
Artigo em Inglês | MEDLINE | ID: mdl-26653411

RESUMO

Since the first implant in 1980, implantable cardioverter defibrillator (ICD) technology has progressed rapidly. Modern ICD's have hundreds of programmable options with the general goal of preventing inappropriate shocks and providing shocks for truly life threatening symptomatic ventricular arrhythmias. New studies on ICD programming have shown the benefits of prolonged detection intervals in reaching this goal. Anti-tachycardia pacing (ATP) therapy has become an important adjunct to defibrillator shocks. Remote monitoring technologies have surfaced which have been shown to identify arrhythmias and problems with the device in an expedient fashion. The subcutaneous ICD offers the advantage of avoiding intravascular leads and their inherent risks. Lastly, the current understanding of the effects of MRI in ICD patients has advanced creating new opportunities to provide MRI safely to such patients.


Assuntos
Arritmias Cardíacas/terapia , Desfibriladores Implantáveis , Taquicardia Ventricular/terapia , Desfibriladores Implantáveis/efeitos adversos , Cardioversão Elétrica , Humanos , Taquicardia Ventricular/fisiopatologia
15.
J Cardiovasc Electrophysiol ; 26(8): 879-884, 2015 Aug.
Artigo em Inglês | MEDLINE | ID: mdl-25917337

RESUMO

BACKGROUND: There are limited data whether history of atrial tachyarrhythmia (AT) modifies the risk of inappropriate ICD therapy, or the efficacy of novel ICD programming to reduce inappropriate ICD therapy events. METHODS: In MADIT-RIT, we investigated the effects of novel ICD programming with high-rate cut-off VT zone ≥ 200 bpm (arm B), or 60-second delayed therapy in the VT zone 170-199 bpm (arm C), compared to conventional programming VT zone>170 bpm (arm A) on first inappropriate ICD therapy in those with or those without AT prior to enrollment. RESULTS: In patients with prior AT (n = 203, 14%) there was a higher risk of inappropriate ICD therapy (HR = 2.10, 95% CI: 1.38-3.20, P < 0.001), and inappropriate ICD shock (HR = 2.56, 95% CI: 1.38-4.74, P = 0.003) compared to those with no prior AT. The effects of innovative programming to reduce inappropriate ICD therapy with either high-rate cut-off or delayed VT therapy were similar in patients with prior AT (arm B vs. A HR = 0.11, P < 0.001, arm C vs. A HR = 0.17, P < 0.001), and also in patients without prior AT before enrollment (arm B vs. A HR = 0.15, P < 0.001, arm C vs. A HR = 0.24, P < 0.001, interaction P-values >0.10 for all). CONCLUSIONS: Novel ICD programming with a high-rate cut-off or delayed therapy is equally beneficial to reduce inappropriate ICD therapy in patients with or without prior AT, despite the lower risk of inappropriate ICD therapy in patients without prior AT.


Assuntos
Fibrilação Atrial/complicações , Flutter Atrial/complicações , Morte Súbita Cardíaca/prevenção & controle , Desfibriladores Implantáveis , Cardioversão Elétrica/instrumentação , Frequência Cardíaca , Prevenção Primária/instrumentação , Falha de Prótese , Taquicardia Ventricular/terapia , Idoso , Fibrilação Atrial/diagnóstico , Fibrilação Atrial/fisiopatologia , Flutter Atrial/diagnóstico , Flutter Atrial/fisiopatologia , Canadá , Morte Súbita Cardíaca/etiologia , Cardioversão Elétrica/efeitos adversos , Europa (Continente) , Feminino , Humanos , Israel , Japão , Masculino , Pessoa de Meia-Idade , Prevenção Primária/métodos , Desenho de Prótese , Fatores de Risco , Taquicardia Ventricular/complicações , Taquicardia Ventricular/diagnóstico , Taquicardia Ventricular/fisiopatologia , Fatores de Tempo , Resultado do Tratamento , Estados Unidos
16.
Europace ; 16(6): 935-8, 2014 Jun.
Artigo em Inglês | MEDLINE | ID: mdl-24864305

RESUMO

The purpose of this European Heart Rhythm Association (EHRA) survey was to examine the current practice on the choice of implantable cardioverter-defibrillator (ICD) type, use of defibrillation testing, and ICD programming for detection and therapy of ventricular arrhythmias. In accordance with recent guidelines and the results of observational studies, the majority of EHRA research network centres reported a high utilization rate of dual-chamber ICDs in the presence of symptomatic and asymptomatic sinus node dysfunction, biventricular ICD in high-degree atrioventricular block and QRS duration <120 ms, and a limited use of defibrillation testing either in primary and secondary prevention settings. Activation of the long ventricular tachycardia (VT) detection window, slow VT zone, antitachycardia pacing before shock for slow and fast VT, and atrial tachyarrhythmia discrimination were considered useful in ICD programming for the majority of patients.


Assuntos
Desfibriladores Implantáveis/estatística & dados numéricos , Análise de Falha de Equipamento/estatística & dados numéricos , Pesquisas sobre Atenção à Saúde , Insuficiência Cardíaca/prevenção & controle , Padrões de Prática Médica/estatística & dados numéricos , Terapia Assistida por Computador/estatística & dados numéricos , Europa (Continente)/epidemiologia , Insuficiência Cardíaca/epidemiologia , Humanos
17.
Cardiol Clin ; 32(2): 191-200, 2014 May.
Artigo em Inglês | MEDLINE | ID: mdl-24793796

RESUMO

Sudden cardiac death is a leading cause of death in the United States and Europe. Implantable cardioverter defibrillators (ICDs) are a cornerstone of therapy for patients at risk of first occurrence of ventricular arrhythmia, or secondary prevention in those who have previously suffered cardiac arrest or life-threatening arrhythmias. Despite their efficacy, ICD shocks are associated with significant physical and psychological adverse effects. As technology has progressed, newer device programing methods have allowed for arrhythmia suppression and termination without the need for high-energy defibrillation, thus improving patient satisfaction, health, and outcomes.


Assuntos
Algoritmos , Morte Súbita Cardíaca/prevenção & controle , Desfibriladores Implantáveis/efeitos adversos , Taquicardia/terapia , Antiarrítmicos/uso terapêutico , Estimulação Cardíaca Artificial , Ablação por Cateter/métodos , Terapia Combinada , Cardioversão Elétrica/efeitos adversos , Traumatismos por Eletricidade/prevenção & controle , Humanos , Satisfação do Paciente , Desenho de Prótese , Falha de Prótese , Qualidade de Vida , Resultado do Tratamento , Procedimentos Desnecessários
18.
Indian Heart J ; 66 Suppl 1: S88-100, 2014.
Artigo em Inglês | MEDLINE | ID: mdl-24568835

RESUMO

BACKGROUND: Appropriate ICD programming is the key to prevent inappropriate shock delivery, that is closely associated to a negative patients' outcome. METHODS: Review of the literature on ICD therapy to generate ICD programmings that can be applied to the broad population of ICD and CRT-D carriers. RESULTS: Arrhythmia detection should occur with a detection time ranging 9″-12″ in the VF zone, and 15″-60″ in the VT zone. Discriminator should be applied at least up to 200 bpm. ATP therapy is applied to all VTs up to 250 bpm, with a success rate of 70%. Inappropriate shocks should occur in <3.6% of patients. CONCLUSION: Tailored ICD programming can be achieved following evidence from large ICD trials. Pre-defined settings that are saved on the programmer and that can be uploaded at device implantation help to ensure optimal programming and to avoid random errors.


Assuntos
Arritmias Cardíacas/terapia , Desfibriladores Implantáveis , Morte Súbita Cardíaca/prevenção & controle , Desenho de Equipamento , Humanos
19.
Heart Rhythm ; 11(3): 426-34, 2014 Mar.
Artigo em Inglês | MEDLINE | ID: mdl-24321235

RESUMO

BACKGROUND: Inappropriate shocks (IASs) complicate implantable cardioverter-defibrillator (ICD) therapy. The management of IASs in patients with a subcutaneous ICD (S-ICD) differs from that in patients with a conventional ICD because of different sensing algorithms and programming options. OBJECTIVE: To describe the management of IASs in patients with an S-ICD. METHODS: Patients were implanted with an S-ICD between February 2009 and July 2012. The prevalence data and clinical determinants of IASs were prospectively collected. In the case of T-wave oversensing (TWOS), an exercise test was performed, and all possible sensing vectors were screened for TWOS. The absence of TWOS defined a suitable vector. RESULTS: Eleven of 69 patients (54% men; mean age 39 ± 14 years; 73% primary prevention) received IASs after 8.9 ± 10 months of implantation (10.8% annual incidence rate). In 8 cases, TWOS caused IASs. Seven of these IASs occurred during exercise and 1 during atrial fibrillation with a high ventricular rate. To manage TWOS, in 7 of 8 patients the sensing vector was changed and in 5 of 8 patients the (un)conditional zone was changed. Hereafter, IASs recurred in 3 of 8 patients, in 2 because of programming errors. Hence, after reprogramming, we observed no IASs in 87.5% of the patients with TWOS during a follow-up of 14.1 ± 13 months. CONCLUSIONS: IASs due to TWOS in the S-ICD can be managed by reprogramming the sensing vector and/or the therapy zones of the device using a template acquired during exercise. Exercise-optimized programming can reduce future IASs, and standard exercise testing shortly after the implantation of an S-ICD may be considered in patients at an increased risk for TWOS.


Assuntos
Arritmias Cardíacas/terapia , Desfibriladores Implantáveis , Cardioversão Elétrica/efeitos adversos , Cardioversão Elétrica/instrumentação , Falha de Prótese , Adulto , Algoritmos , Eletrocardiografia , Feminino , Humanos , Masculino , Estudos Prospectivos , Desenho de Prótese , Fatores de Risco
SELEÇÃO DE REFERÊNCIAS