1.
Dent Mater ; 40(1): 80-89, 2024 Jan.
Artigo em Inglês | MEDLINE | ID: mdl-37919111

RESUMO

OBJECTIVES: To evaluate the effect of mono and multi-wave light-curing units (LCUs) on the Knoop hardness of resin-based composites (RBC) that use different photoinitiators. METHODS: Central incisor-shaped specimens 12 mm long, 9 mm wide, and 1.5 mm thick were made from 2 RBCs that use different photoinitiators: Tetric N-Ceram (Ivoclar Vivadent) - and Vittra APS (FGM), both A2E shade. They were light-cured with 4 different LCUs: two claimed to be multi-wave - VALO Grand (Ultradent) and Emitter Now Duo (Schuster); and two were monowave - Radii Xpert (SDI) and Elipar DeepCure-L (3 M Oral Care) using 2 different light exposure protocols: one 40 s exposure centered over the specimen; and two 20 s light exposures that delivered light from two positions to better cover the entire tooth. 16 groups with 10 specimens in each group were made. The Knoop hardness (KH, kg/mm2) was measured at the top and bottom of the specimen in the center and at the cervical, incisal, mesial, and distal peripheral regions. The active tip diameters (mm) and spectral radiant powers (mW/nm) of the LCUs were measured with and without the interposition of the RBC, as well as the radiant exposure beam profiles (J/cm²) delivered to the top of the RBCs. The data was analyzed using Three-way ANOVA and Tukey's tests (α = 0.05). RESULTS: The VALO Grand (1029 mW) emitted twice the power of the Radii Xpert (500 mW). The KH values of VI and TN resin composite specimens were significantly affected by the LCU used (p < .001), the measurement location (p < .001), and the surface of the specimen (p < .001). LCUs with wider tip diameters produced greater Knoop hardness values at the peripheries of the 12 mm of long, 9 mm wide specimens. In general, the VALO Grand produced the highest KH values, followed by Elipar DeepCure-L, then by Radii Xpert. The Emitter Now Duo LCU produced the lowest values. Exposing the veneers from two locations reduced the differences between the LCUs and the effect of the measurement location. Only the VALO Grand could fully cover the composite veneer with light when the two locations were used. SIGNIFICANCE: The light tip must cover the entire restoration to photocure the RBC beneath the light tip.


Assuntos
Lâmpadas de Polimerização Dentária , Cura Luminosa de Adesivos Dentários , Dureza , Teste de Materiais , Resinas Compostas , Materiais Dentários , Polimerização
2.
Braz. dent. j ; Braz. dent. j;28(3): 362-371, May-June 2017. tab, graf
Artigo em Inglês | LILACS | ID: biblio-888649

RESUMO

Abstract This study measured the radiant power (mW), irradiance (mW/cm2) and emission spectra (mW/cm2/nm) of 22 new, or almost new, light curing units (LCUs): - Alt Lux II, BioLux Standard, Bluephase G2, Curing Light XL 3000, Demetron LC, DX Turbo LED 1200, EC450, EC500, Emitter C, Emitter D, KON-LUX, LED 3M ESPE, Led Lux II, Optilight Color, Optilight Max, Optilux 501, Poly Wireless, Radii cal, Radii plus, TL-01, VALO Cordless. These LCUs were either monowave or multiple peak light emitting diode (LED) units or quartz-tungsten-halogen LCUs used in anterior and posterior teeth. The radiant power emitted by the LCUs was measured by a laboratory grade laser power meter. The tip area (cm²) of the LCUs was measured and used to calculate the irradiance from the measured radiant power source. The MARC-Patient Simulator (MARC-PS) with a laboratory grade spectrometer (USB4000, Ocean Optics) was used to measure the irradiance and emission spectrum from each LCU three times at the sensor located on the facial of the maxillary central incisors and then separately at the occlusal of a maxillary second molar. The minimum acceptable irradiance level was set as 500 mW/cm2. Irradiance data was analyzed using two-way ANOVA and the radiant power data was analyzed by one-way ANOVA followed by Tukey test (a=0.05). In general, the irradiance was reduced at the molar tooth for most LCUs. Only the Valo, Bluephase G2 and Radii Plus delivered an irradiance similar to the anterior and posterior sensors greater than 500 mW/cm2. KON-LUX, Altlux II, Biolux Standard, TL-01, Optilux 501, DX Turbo LED 1200 LCUs delivered lower irradiance values than the recommended one used in molar region, KON-LUX and Altlux II LCUs used at the maxillary incisors. Bluephase G2 and Optilight Max delivered the highest radiant power and KON-LUX, Altlux II and Biolux Standard delivered the lowest power. The emission spectrum from the various monowave LED LCUs varied greatly. The multi-peak LCUs delivered similar emission spectra to both sensors.


Resumo Este estudo mediu a potência (mW), irradiância (mW/cm2) e espectro da luz (mW/cm2/nm) emitida por 22 fontes de luz (Alt Lux II, BioLux Standard, Bluephase G2, Curing Light XL 3000, Demetron LC, DX Turbo LED 1200, EC450, EC500, Emitter C, Emitter D, KON-LUX, LED 3M ESPE, Led Lux II, Optilight Color, Optilight Max, Optilux 501, Poly Wireless, Radii cal, Radii plus, TL-01, VALO Cordless) disponíveis comercialmente. A potência emitida pelas fontes de luz foi medida usando um medidor laboratorial de potencia com grade a laser. A área (cm²) da ponta ativa efetiva das fontes de luz foi medida com paquímetro digital e utilizada para calcular a irradiância emitida. O simulador de paciente-MARC (MARC - PS) com espectrómetro (USB4000, Ocean Optics) foi usado para medir a irradiância e o espectro de luz emitida por cada fonte de luz na região anterior e posterior. Esta medição foi repetida por três vezes em dois sensores localizados na região anterior e posterior da arcada dentária. Os dados de irradiância foram analisados utilizando análise de variância em dois fatores, e os dados de potência foram analisados com análise de variância em fator único seguido pelo teste de Tukey (a=0,05). As fontes de luz Valo, Bluephase G2, Radii Plus emitiram irradiância semelhante tanto na região anterior como posterior com valores superiores ao mínimo de 500 mW/cm2. Seis fontes de luz emitiram irradiância menor que o recomendado (500 mW/cm2) quando usadas na região posterior: Kon-lux, Altlux II, Biolux Standard TL-01, Optilux 501, DX Turbo LED 1200 e duas quando usadas na região anterior: Kon-lux e Altlux II LCUs. As fontes Bluephase G2, Optilight Max emitiram os maiores valores de potência e as fontes de luz Altlux II e Biolux Standard emitiram os menores valores de potência. O espectro de luz das fontes LED de espectro único variou de forma evidente entre as fontes. As fontes LED multi pico de espectro emitiram espectros de luz similar para ambos os sensores. A fotoativação na região posterior tende a reduzir substancialmente a irradiância da maioria das fontes de luzes testadas.


Assuntos
Humanos , Lâmpadas de Polimerização Dentária , Restauração Dentária Permanente , Lasers , Simulação de Paciente , Análise Espectral/instrumentação , Dente/efeitos da radiação
3.
Braz. dent. j ; Braz. dent. j;28(1): 9-15, Jan.-Feb. 2017. tab, graf
Artigo em Inglês | LILACS | ID: biblio-839109

RESUMO

Abstract The high irradiance and the different emission spectra from contemporary light curing units (LCU) may cause ocular damage. This study evaluated the ability of 15 eye protection filters: 2 glasses, 1 paddle design, and 12 dedicated filters to block out harmful light from a monowave (HP-3M ESPE) and a broad-spectrum (Valo, Ultradent) LED LCU. Using the anterior sensor in the MARC-Patient Simulator (BlueLight Analytics) the irradiance that was delivered through different eye protection filters was measured three times. The LCUs delivered a similar irradiance to the top of the filter. The mean values of the light that passed through the filters as percent of the original irradiance were analyzed using two-way ANOVA followed by Tukey test (a= 0.05). The emission spectra from the LCUs and through the filters were also obtained. Two-way ANOVA showed that the interaction between protective filters and LCUs significantly influenced the amount of light transmitted (p< 0.001). Tukey test showed that the amount of light transmitted through the protective filters when using the HP-3M-ESPE was significantly greater compared to when using the Valo, irrespective of the protective filter tested. When using the HP-3M-ESPE, the Glasses filter allowed significantly more light through, followed by XL 3000, ORTUS, Google Professional, Gnatus filters. The Valo filter was the most effective at blocking out the harmful light. Some protective filters were less effective at blocking the lower wavelengths of light (<420 nm). However, even in the worst scenario, the filters were able to block at least 97% of the irradiance.


Resumo A alta irradiância e diferentes espectros de luz emitidos por aparelhos fotopolimerizadores (Fp) podem causar danos oculares. Este estudo avaliou a capacidade de 15 filtros de proteção ocular em bloquear a luz prejudicial de um Fp convencional (HP-3M ESPE) e outro de largo espectro (Valo, Ultradent). Utilizando sensor anterior do equioamento MARC-Patient Simulator (BlueLight Analytics inc.) a irradiância que passou através dos diferentes filtros protetores foi mensuradas três vezes. Os valores médios da irradiância que passaram pelos filtros foram analisados usando Análise de variância fatorial e pelo teste de Tukey (a= 0.05). O espetro emitido dos Fps através dos filtros também foi obtido. A análise de variância mostrou que a interação entre os filtros protetores e Fps influenciou significantemente a quantidade de luz transmitida (p<0,001). O teste de Tukey mostrou que a quantidade que luz transmitida através dos protetores oculares quando usado o HP-3M ESPE foi significantemente maior quando comparado aos valores para o Valo, independentemente do filtro testado. Quando foi utilizado a fonte de luz HP-3M ESPE, o filtro de proteção ocular permitiu significativamente maior passagem de luz, seguido por XL 3000, ORTUS, Google Professional, e pelo filtro Gnatus. O filtro do Valo foi o mais eficiente ao bloquear a luz prejudicial. Alguns filtros foram menos eficazes ao bloquear menores comprimentos de onde (<420 nm). No entanto, mesmo no pior cenário dos resultados deste estudo, os filtros foram capazes de bloquear ao menos 97% da irradiância emitida pelas fontes de luz testadas.


Assuntos
Humanos , Lâmpadas de Polimerização Dentária/efeitos adversos , Dispositivos de Proteção dos Olhos/normas , Análise de Variância , Óptica e Fotônica
4.
Restor Dent Endod ; 41(4): 271-277, 2016 Nov.
Artigo em Inglês | MEDLINE | ID: mdl-27847748

RESUMO

OBJECTIVES: In this study, we evaluated the influence of different radiant exposures provided by single-peak and polywave light-curing units (LCUs) on the degree of conversion (DC) and the mechanical properties of resin cements. MATERIALS AND METHODS: Six experimental groups were established for each cement (RelyX ARC, 3M ESPE; LuxaCore Dual, Ivoclar Vivadent; Variolink, DMG), according to the different radiant exposures (5, 10, and 20 J/cm2) and two LCUs (single-peak and polywave). The specimens were made (7 mm in length × 2 mm in width × 1 mm in height) using silicone molds. After 24 hours of preparation, DC measurement was performed using Fourier transform infrared spectrometry. The same specimens were used for the evaluation of mechanical properties (flexural strength, FS; elastic modulus, E) by a three-point bending test. Data were assessed for normality, after which two-way analysis of variance (ANOVA) and post hoc Tukey's test were performed. RESULTS: No properties of the Variolink cement were influenced by any of the considered experimental conditions. In the case of the RelyX ARC cement, DC was higher when polywave LCU was used; FS and E were not influenced by the conditions evaluated. The LuxaCore cement showed greater sensitivity to the different protocols. CONCLUSIONS: On the basis of these results, both the spectrum of light emitted and the radiant exposure used could affect the properties of resin cements. However, the influence was material-dependent.

5.
Odontol. vital ; jun. 2016.
Artigo em Espanhol | LILACS-Express | LILACS | ID: biblio-1506808

RESUMO

Se sabe que durante el fotocurado de los materiales dentales se libera calor, ya sea por parte de la fuente emisora de luz o por la reacción de polimerización del compuesto utilizado. Este aumento de temperatura se transmite a los tejidos, el cual puede repercutir en ellos. Debido a esto, se realiza un estudio comparativo de la temperatura alcanzada durante la obturación y reconstrucción de piezas dentales in vitro, utilizando lámparas de fotocurado de tipo cuarzo-tungsteno-halógeno (QTH) , LED y lámparas conformadas por 4 LED, y se analiza sus programaciones: modo rampa, modo intermitente y normal; y se proponen pautas clínicas para reducir los riesgos térmicos potenciales a la pulpa y tejidos de soporte.


It is known that the heat is realeased during light-curing of dental materials, either by the light emitting source or by the polymerization reaction of the compound used. This temperature increase is transmitted to tissues, which may have an effect on them. Due to this, a comparative study of the temperature reached during the sealing and reconstruction of teeth In vitro was performed, using light-curing lamps from the type quartztungsten- halogen (QTH), LED and lamps formed by 4 LEDs, analyzing their programming: Ramp Mode, Intermittent Mode, and normal: and clinical guidelines are proposed in order to reduce potential thermal risks to the pulp and supporting tissues

6.
J Dent ; 43(12): 1565-72, 2015 Dec.
Artigo em Inglês | MEDLINE | ID: mdl-26343422

RESUMO

OBJECTIVES: To evaluate the effect of photoinitiators and reducing agents on cure efficiency and color stability of resin-based composites using different LED wavelengths. METHODS: Model resin-based composites were associated with diphenyl(2,4,6-trimethylbenzoyl) phosphine oxide (TPO), phenylbis(2,4,6-trimethylbenzoyl) phosphine oxide (BAPO) or camphorquinone (CQ) associated with 2-(dimethylamino) ethyl methacrylate (DMAEMA), ethyl 4-(dimethyamino) benzoate (EDMAB) or 4-(N,N-dimethylamino) phenethyl alcohol (DMPOH). A narrow (Smartlite, Dentisply) and a broad spectrum (Bluephase G2, Ivoclar Vivadent) LEDs were used for photo-activation (20 J/cm(2)). Fourier transform infrared spectroscopy (FT-IR) was used to evaluate the cure efficiency for each composite, and CIELab parameters to evaluated color stability (ΔE00) after aging. The UV-vis absorption spectrophotometric analysis of each photoinitiator and reducing agent was determined. Data were analyzed using two-way ANOVA and Tukey's test for multiple comparisons (α=0.05). RESULTS: Higher cure efficiency was found for type-I photoinitiators photo-activated with a broad spectrum light, and for CQ-systems with a narrow band spectrum light, except when combined with an aliphatic amine (DMAEMA). Also, when combined with aromatic amines (EDMAB and DMPOH), similar cure efficiency with both wavelength LEDs was found. TPO had no cure efficiency when light-cured exclusively with a blue narrowband spectrum. CQ-systems presented higher color stability than type-I photoinitiators, especially when combined with DMPOH. CONCLUSIONS: After aging, CQ-based composites became more yellow and BAPO and TPO lighter and less yellow. However, CQ-systems presented higher color stability than type-I photoinitiators, as BAPO- and TPO-, despite their higher cure efficiency when photo-activated with corresponding wavelength range. CLINICAL SIGNIFICANCE: Color matching is initially important, but color change over time will be one of the major reasons for replacing esthetic restorations; despite the less yellowing of these alternative photoinitiators, camphorquinone presented higher color stability.


Assuntos
Bis-Fenol A-Glicidil Metacrilato/química , Cor , Lâmpadas de Polimerização Dentária/classificação , Cura Luminosa de Adesivos Dentários/métodos , Fotoiniciadores Dentários/química , Substâncias Redutoras/química , Resinas Acrílicas/química , Cânfora/análogos & derivados , Cânfora/química , Resinas Compostas/química , Luz , Teste de Materiais , Metacrilatos/química , Fosfinas/química , Poliuretanos/química , Espectroscopia de Infravermelho com Transformada de Fourier
7.
Rev. clín. pesq. odontol. (Impr.) ; 5(2): 111-119, maio-ago. 2009. tab, ilus
Artigo em Inglês | LILACS, BBO - Odontologia | ID: lil-617410

RESUMO

OBJECTIVE: To investigate the effect of light curing units and storage time on the degree of conversion of fissure sealants. MATERIAL AND METHOD: Three different fissure sealants were used: a resin-based pit and fissure sealant (Clinpro SealantTM), a polyacid modified resin composite (Dyract SealTM), and a resin modified glass ionomer cement (VitremerTM). Each fissure sealant was polymerized with a halogen, high intensity quartz tungsten halogen, LED. Half of the specimens were tested to assess the degree of conversion immediately after polymerization and the others after one month’s storage in artificial saliva. The degree of conversion was measured by FTIR. The data were analyzed using two way ANOVA. RESULTS: Statistical analysis of Dyract SealTM showed significant differences between the DC values of samples tested immediately and again after one month’s storage (p < 0.05). The DC values of ClinproTM increased after the storage period (p < 0.05). The DC of VitremerTM showed no change after storage (p > 0.05). Results of the curing units testing showed no significant differences among the DC of samples polymerized with LED, HQTH, OTH tested immediately and again after one month for Dyract SealTM, VitremerTM and Clinpro SealantTM (p > 0.05). CONCLUSION: Light curing units had no effect on the degree of conversion of the fissure sealants. But storage after initial light activation and polymerization for one month in artificial saliva can affects on DC of fissure sealants.


OBJETIVO: Investigar o efeito de unidades de polimerização por luz no grau de conversão dos selantes de fissuras. MATERIAL E MÉTODO: Três diferentes selantes foram utilizados: um selante resinoso de fissuras (Clinpro SealantTM, uma resina composta poliácida modificada (DyractSealTM) cimento de ionômero de vidro modificado (VitremerTM). Cada selante de fissura foi polimerizado por LED halógeno de quartzo/tungstênio. Metade dos espécimes foi testada para assegurar o grau de conversão imediatamente após polimerização e outros após um mês de armazenamento em saliva artificial. O grau de conversão foi medido por FTIR. Os resultados foram analisados por meio do ANOVA. RESULTADOS: A análise estatística do Dyract SealTMdemonstrou diferenças estatisticamente significantes entre os valores DC dos espécimes testados imediatamente e novamente após um mês de armazenamento (p < 0,05). Os valores do ClimproTM aumentaram após o período de armazenamento (p < 0.05). O DC do VitremerTM não demonstraram alterações após o armazenamento (p > 0.05). Os resultados das unidades de fotopolimerização não mostraram diferenças significantes ente DC de amostras polimerizadas com LED, HQTH, O TH testados imediatamente e novamente após um mês para o Diract SealTM, VitremerTM e Climpro SealantTM. CONCLUSÃO: Unidades de fotopolimerização não tiveram efeito no grau de conversão dos selantes de fissuras. Porém, armazenamento após ativação inicial e polimerização por um mês em saliva artificial pode afetar o DC dos selantes de fissuras.


Assuntos
Cimentos de Ionômeros de Vidro , Resinas Compostas/efeitos da radiação , Selantes de Fossas e Fissuras/efeitos da radiação , Análise de Variância , Polímeros/efeitos da radiação , Saliva Artificial , Espectroscopia de Infravermelho com Transformada de Fourier , Fatores de Tempo
8.
Braz. dent. j ; Braz. dent. j;20(5): 410-413, 2009. ilus, tab
Artigo em Inglês | LILACS | ID: lil-537548

RESUMO

The aim of this study was to evaluate the Knoop hardness after 15 min and 24 h of different shades of a dual-cured resin-based cement after indirect photoactivation (ceramic restoration) with 2 light-curing units (LCUs). The resin cement Variolink II (Ivoclar Vivadent) shade XL, A2, A3 and opaque were mixed with the catalyst paste and inserted into a black Teflon mold (5 mm diameter x 1 mm high). A transparent strip was placed over the mold and a ceramic disc (Duceram Plus, shade A3) was positioned over the resin cement. Light-activation was performed through the ceramic for 40 s using quartz-tungsten-halogen (QTH) (XL 2500; 3M ESPE) or light-emitting diode (LED) (Ultrablue Is, DMC) LCUs with power density of 615 and 610 mW/cm2, respectively. The Koop hardness was measured using a microhardness tester HMV 2 (Shimadzu) after 15 min or 24 h. Four indentations were made in each specimen. Data were subjected to ANOVA and Tukey's test (a=0.05). The QTH LCU provided significantly higher (p<0.05) KHN values than the LED LCU. When the post-cure times were compared for the same shade, QTH and LED at 24 h provided significantly higher (p<0.05) KHN values than at 15 min. It may be concluded that the Knoop hardness was generally dependent on the LCU and post-cure time. The opaque shade of the resin cement showed lower Knoop hardness than the other shades for both LCUs and post-cure times.


Este estudo avaliou a dureza Knoop após 15 min e 24 h de diferentes cores do cimento resinoso dual após fotoativação indireta com 2 fontes de luz. O cimento resinoso Variolink II (Ivoclar Vivadent) nas cores XL, A2, A3 e opaco foram misturados com a pasta catalisadora e inseridos dentro de um molde de teflon preto com (5 mm de diâmetro x 1 mm de espessura). A tira de poliéster transparente foi inserida sobre o molde e um disco da cerâmica (Duceram Plus, cor A3) foi posicionada sobre o cimento resinoso. A fotoativação foi realizada sobre o disco de cerâmica por 40 s usando a lâmpada de luz halógena (XL 2500, 3M ESPE) ou luz emitida por diodo (LED) com intensidade de luz de 615 e 610 mW/cm2. A dureza Knoop foi medida usando o microdurômetro HMV 2 (Shimadzu) com carga de 50 g por 15 s, após 15 min e 24 h. Quatro indentações foram feitas em cada amostra. Os dados foram submetidos à Análise de Variância e ao teste de Tukey (a=5 por cento). A fonte de luz halógena promoveu valores de dureza significantemente maiores (p<0,05) que o LED. Quando os tempos pós-polimerização foram comparados para a mesma cor, a luz halógena e o LED promoveram valores de dureza significantemente maiores (p<0,05) em 24 h do que aos 15 min. Pode-se concluir que, de maneira geral, a dureza Knoop foi dependente da fonte de luz e do tempo pós-polimerização. A cor opaca do cimento resinoso mostrou menores valores de dureza Knoop que as outras cores para ambas as fontes de luz e tempos pós-polimerização.


Assuntos
Lâmpadas de Polimerização Dentária , Porcelana Dentária , Cura Luminosa de Adesivos Dentários , Cimentos de Resina , Análise de Variância , Análise do Estresse Dentário , Dureza , Teste de Materiais , Cimentos de Resina/química , Cimentos de Resina/efeitos da radiação , Estatísticas não Paramétricas , Fatores de Tempo
9.
Rev. odontol. UNESP ; 35(4): 293-298, out.-dez. 2006. ilus, tab
Artigo em Português | LILACS, BBO - Odontologia | ID: biblio-858468

RESUMO

A proposta do trabalho foi avaliar o grau de microinfiltração marginal do sistema adesivo Single Bond (3M ESPE) por meio de dois diferentes tipos de fotopolimerizadores. Foram utilizados dez pré-molares humanos e confeccionadas cavidades do tipo classe II, sendo uma ocluso-mesial e outra ocluso-distal, com término oclusal em esmalte e término cervical localizado em cemento, em seguida, restauradas com Single Bond (3M ESPE) e resina composta Filtek Z250 (3M ESPE) cor A2. Os dentes foram divididos em dois grupos: Grupo 1 û fotopolimerizado com o aparelho de luz halógena Optilight (Gnatus); Grupo 2 û fotopolimerizado com o aparelho sem fio LED Radii (SDI). Os dentes e os forames apicais foram isolados, deixando exposta à margem de 1,0 mm ao redor da restauração, sendo em seguida termociclados 250 vezes (±5°C e ±55°C). Após, os corpos-de-prova foram mergulhados em solução de fucsina básica a 0,5 porcento por 48 horas para posterior avaliação obedecendo aos escores: grau 0 û ausência de infiltração; grau 1 û infiltração até 1/3 da parede; grau 2 û infiltração entre 1/3 a 2/3 da parede e grau 3 û infiltração mais do que 2/3 da parede. Os escores para dentina foram: G1: 0 = 100 porcento; G2: 0 = 70 porcento; 1 = 10 porcento; 2 = 20 porcento. Para esmalte: G1: 0 = 90 porcento; 1 = 10 porcento; G2: 0 = 50 porcento; 2 = 20 porcento; G3 = 30 porcento. Os resultados foram submetidos à análise estatística de Mann-Whitney (p < 0,05), demonstrando não haver diferença estatisticamente significante entre os grupos. Concluiu-se que a microinfiltração marginal não foi influenciada negativamente quando foi empregada a unidade fotopolimerizadora LED


The purpose of this study was to evaluate the microleakage of the bonding system Single Bond (3M ESPE) when used 2 differents light-curing units. Class II cavities were prepared in 10 human premolars. After, the cavities prepared were treated with the bonding agent Single Bond (3M ESPE) and filled with Filtek Z250 (3M ESPE) composite resin, color A2. The teeth were divided into 2 groups: G1 ­ the restorations were light-cured with a conventional halogen-based light-curing unit (Gnatus); G2 ­ the restorations were cured with a LED cordless light-curing unit Radii (SDI). The specimes were thermo-cycled 250 cycles (± 5 °C and ± 55 °C) prior to immersion in 0.5% basic fuchsin for 48 hours for posterior evaluation. The scores in dentin were: G1: 0 = 100%; G2: 0 = 70%; 1 = 10%; 2 = 20%. In the enamel: G1: 0 = 90%; 1 = 10%; G2: 0 = 50%;­ 2 = 20%; G3 = 30%. The data were subjected to statistical analysis using Mann-Whitney test (p < 0.05) and showed no significant statistical difference between the groups. In conclusion the LED light source no influenced the microleakage


Assuntos
Humanos , Infiltração Dentária , Restauração Dentária Permanente , Cura Luminosa de Adesivos Dentários , Lâmpadas de Polimerização Dentária , Polimerização , Resinas Compostas , Dente Pré-Molar , Adesivos Dentinários , Estatísticas não Paramétricas