1.
Clin Biochem
; 19(3): 175-8, 1986 Jun.
Artigo
em Inglês
| MEDLINE
| ID: mdl-3089643
RESUMO
An automated enzymatic method is described for the determination of Cu,Zn-superoxide dismutase (SOD) in plasma or erythrocytes using the xanthine-xanthine oxidase and cytochrome C coupled assay. This method was adapted to an Abbott ABA-200 discrete analyzer. Coefficients of variation for within-run and day-to-day analyses were less than 5%. Only 2.5 muL of serum or erythrocyte extract is required so a capillary tube sample of blood (70 muL) is sufficient for the assay. Recovery of added SOD ranged from 92 to 101%. The method reported here is practical for use in a clinical chemistry laboratory for monitoring changes in this enzyme, which is a sensitive early indicator of alterations in copper status.