Método espectrofotométrico na região UV a 205 nm para determinação de íon nitrato em água: validação / Spectrophotometric method in UV region at 205 nm for determination of nitrate ion in water: validation
Scorsafava, Maria Anita; Souza, Arlete de; Alves, Gisele Letícia.
Affiliation
  • Scorsafava, Maria Anita; Instituto Adolfo Lutz. Centro de Contaminante. Núcleo de Águas e Embalagens. São Paulo. BR
  • Souza, Arlete de; Instituto Adolfo Lutz. Centro de Contaminante. Núcleo de Águas e Embalagens. São Paulo. BR
  • Alves, Gisele Letícia; Instituto Adolfo Lutz. Centro de Contaminante. Núcleo de Águas e Embalagens. São Paulo. BR
Rev. Inst. Adolfo Lutz (Online) ; 76: e1731, 2017. tab, graf
Article in Pt | LILACS, VETINDEX | ID: biblio-1489561
Responsible library: BR68.1
RESUMO
Este estudo teve por objetivo validar o método espectrofotométrico na região ultravioleta(205 nm) para determinação de nitrato em água para consumo humano e de diálise, bemcomo calcular sua incerteza associada. A linearidade, seletividade, precisão, exatidão, limitede detecção (LD) e limite de quantificação (LQ) foram avaliados de acordo com o documentode orientação INMETRO DOQ-CGCRE-008. Os resultados obtidos utilizando água deabastecimento indicaram que a matriz não tem efeito significativo sobre a seletividade dométodo. A exatidão e a repetitividade foram avaliadas com três níveis de concentração doanalito na água de abastecimento, na água de diálise e em material de referência certificado,apresentando resultados satisfatórios para todas as matrizes utilizadas. Os limites de detecção equantificação foram de 0,074 e 0,248 mg/L de nitrato, respectivamente. O método mostrou-seadequado para determinação de nitrato em água de abastecimento e em água de diálise.
ABSTRACT
This study validated the spectrophotometric method in ultraviolet region (205 nm) fordetermining nitrate in water used for human consumption and dialysis procedure and tocalculate its associated uncertainty. The linearity, selectivity, precision, accuracy, limit ofdetection (LD) and limit of quantification (LQ) were evaluated according to the INMETRODOQ-CGCRE-008. The results obtained using water supply indicated that the matrix has nosignificant effect on the selectivity of the method. Accuracy and repeatability were evaluatedusing three different levels of analyte concentration in the supply water, in the dialysis waterand in the certified reference material, showing satisfactory results at all matrices used.The limits of detection and of quantification were 0.074 and 0.248 mg/L of nitrate, respectively.The method proved to be adequate for determining nitrate in water supply and dialysis water.
Subject(s)
Key words
Full text: 1 Collection: 01-internacional Database: LILACS / VETINDEX Main subject: Spectrophotometry, Ultraviolet / Water Microbiology / Dialysis / Ions / Nitrates Language: Pt Journal: Rev. Inst. Adolfo Lutz (Online) Journal subject: Medicina / Microbiologia / Sa£de P£blica Year: 2017 Document type: Article Affiliation country: Brazil Country of publication: Brazil
Full text: 1 Collection: 01-internacional Database: LILACS / VETINDEX Main subject: Spectrophotometry, Ultraviolet / Water Microbiology / Dialysis / Ions / Nitrates Language: Pt Journal: Rev. Inst. Adolfo Lutz (Online) Journal subject: Medicina / Microbiologia / Sa£de P£blica Year: 2017 Document type: Article Affiliation country: Brazil Country of publication: Brazil