Complicaciones en la implantación coclear / Complications in cochlear implantation
Ramos, A; Charlone, R; Miguel, I de; Valdivielso, A; Cuyas, JM; Pérez, D; Vasallo, JR.
Affiliation
  • Ramos, A; Hospital Universitario Insular de Gran Canaria. Servicio de Otorrinolaringología y Patología Cérvico Facial. Las Palmas de Gran Canaria. España
  • Charlone, R; Hospital Universitario Insular de Gran Canaria. Servicio de Otorrinolaringología y Patología Cérvico Facial. Las Palmas de Gran Canaria. España
  • Miguel, I de; Hospital Universitario Insular de Gran Canaria. Servicio de Microbiología. Las Palmas de Gran Canaria. España
  • Valdivielso, A; Hospital Universitario Insular de Gran Canaria. Servicio de Otorrinolaringología y Patología Cérvico Facial. Las Palmas de Gran Canaria. España
  • Cuyas, JM; Hospital Universitario Insular de Gran Canaria. Servicio de Otorrinolaringología y Patología Cérvico Facial. Las Palmas de Gran Canaria. España
  • Pérez, D; Hospital Universitario Insular de Gran Canaria. Servicio de Otorrinolaringología y Patología Cérvico Facial. Las Palmas de Gran Canaria. España
  • Vasallo, JR; Hospital Universitario Insular de Gran Canaria. Servicio de Otorrinolaringología y Patología Cérvico Facial. Las Palmas de Gran Canaria. España
Acta otorrinolaringol. esp ; 57(3): 122-125, mar. 2006. ilus, tab
Article in Es | IBECS | ID: ibc-045414
Responsible library: ES1.1
Localization: ES1.1 - BNCS
RESUMEN

Objetivo:

El objetivo de este trabajo es mostrar nuestra experiencia en complicaciones quirúrgicas del implante coclear, tanto en la descripción como en el tratamiento de las mismas. Se presentan 346 pacientes, operados por el mismo equipo quirúrgico, sometidos a un control y seguimiento prolongado de tiempo. Material y

métodos:

Se estudió una población de 346 pacientes implantados cocleares, desde febrero de 1993 hasta marzo de 2004. La serie incluye 211 niños y 135 adultos. Se analizan los tipos de complicaciones, su manejo y seguimiento.

Resultados:

El índice de complicaciones fue del 9,8% (n=34).Las complicaciones más frecuentemente halladas fueron las siguientes Complicaciones intraoperatorias fístulas de líquido cefalorraquídeo (LCR) 7; lesiones del nervio facial, 1; excesivo adelgazamiento de pared posterior, 2. Complicaciones postoperatorias inmediatas infección de herida quirúrgica, 4; dolor postoperatorio persistente, 1; acúfenos, 2. Complicaciones postoperatorias tardías mastoiditis, 2; extrusión del receptor estimulador, 1; estimulación del VIIº par, 5; infección tardía de la zona del receptor, 2; fallo técnico del sistema implantado, 7. No se encontró ninguna alteración en la colocación del electrodo ni migración del mismo.

Conclusiones:

La implantación coclear es una cirugía de baja morbilidad. En la serie que se presenta, no hubo ningún caso de meningitis debido a este procedimiento
ABSTRACT
Objetive The objetive of this paper is to present the surgical complications in cochlear implant, in a serie of 346 patients, submitted to surgery by the same surgical team with the subsequent control and follow – up for a long period of time. We show the description and the handling of each complication. Material and

Methods:

A follow – up was carried out on 346 cochlear implant patients by the same surgical team, from February 1993 to March 2004. The complications were assessed and their handling and follow – up is reported. The series includes 211 children and 135 adults.

Results:

The complication rate was 9.8% (n=34). The complications found were Intraoperatory Complications Cerebrospinal Fluid Leaks (CSF), 7; Facial nerve injury, 1; excessive thinning of the posterior wall, 2; inmediate post operatory complications infection of the surgical wound, 4; post operatory persistent pain, 1; Tinnitus, 2; late complications Mastoiditis, 2; extrusion of the stimulating receptor, 1; facial nerve stimulation, 5; late infections of the stimulation receptor, 2; technical failure of the implanted system, 7. Neither alteration or migration was found in the receptor placing.

Conclusions:

The cochlear implant surgery has a low morbility. We found no cases of meningitis due to this procedure
Subject(s)
Search on Google
Collection: National databases / Spain Health context: SDG3 - Target 3.3 End transmission of communicable diseases Health problem: Meningitis Database: IBECS Main subject: Cochlear Implantation Type of study: Etiology study / Observational study / Prognostic study / Risk factors Limits: Adult / Child / Humans Language: Spanish Journal: Acta otorrinolaringol. esp Year: 2006 Document type: Article Institution/Affiliation country: Hospital Universitario Insular de Gran Canaria/España
Search on Google
Collection: National databases / Spain Health context: SDG3 - Target 3.3 End transmission of communicable diseases Health problem: Meningitis Database: IBECS Main subject: Cochlear Implantation Type of study: Etiology study / Observational study / Prognostic study / Risk factors Limits: Adult / Child / Humans Language: Spanish Journal: Acta otorrinolaringol. esp Year: 2006 Document type: Article Institution/Affiliation country: Hospital Universitario Insular de Gran Canaria/España
...